RVT overview

  • 200 practice questions
  • 32 flashcards
  • 5 study guide topics
  • 40 key terms

What you'll study

  • Cerebrovascular20%
  • Peripheral Arterial20%
  • Peripheral Venous20%
  • Abdominal and Visceral Vascular20%
  • Instrumentation, Quality Assurance and Specialized Vascular Testing20%

Registered Vascular Technologist (RVT®) is an ARDMS credential administered by Inteleos for sonographers who perform vascular ultrasound as a core part of their independent practice. Earning it is a two-exam process: candidates must pass the Sonography Principles and Instrumentation (SPI) exam, the foundation exam shared by all ARDMS certifications, plus the Vascular Technology (VT) specialty exam. The VT exam tests knowledge of the anatomy and hemodynamics of blood vessels, and material spans the cerebrovascular circulation, peripheral arterial and peripheral venous systems, abdominal and visceral vessels, and the instrumentation, quality assurance and specialized testing knowledge that underpins every vascular study. Because the SPI is the prerequisite gateway, most candidates sit for it first and then build vascular-specific knowledge; ARDMS requires the SPI to have been passed within five years of the VT attempt. Both exams use year-round application and examination windows, with the SPI priced at $275 USD and the VT priced at $300 USD. RVT certification supports compliance with Intersocietal Accreditation Commission (IAC) Vascular Testing Standards and laboratory accreditation requirements, and it signals validated expertise to employers in hospitals, imaging centers and specialty practices. Preparation means reviewing the published exam topics and study guides, then practicing interpretation of hemodynamic principles rather than memorizing protocols.

Common mistakes

Calculating the ICA/CCA peak systolic ratio using the CCA velocity measured at the bulb rather than at the mid-to-distal common carotid, and reporting a ratio that is falsely low or high.

Candidates memorize 'ratio >3.0 means severe' but do not rehearse where each sample volume belongs; the bulb is a low-velocity, widened segment that drags the denominator down.

What to do instead

Before every practice calculation, state aloud the sampling site for each component (mid-distal CCA, 1-2 cm distal to the bifurcation for ICA), and drill ten ratio calculations where you write the sampling location next to each velocity.

Calling a vessel occluded when no color flow is seen in it, without checking grayscale, power Doppler, and spectral Doppler for a detectable but very low-velocity signal.

Color Doppler is flow- and velocity-dependent; low-velocity or low-volume flow can fall below the wall filter and color scale threshold, and candidates conflate 'no color' with 'no flow'.

What to do instead

For any suspected occlusion, perform a fixed sequence: reduce the color PRF/scale, lower the wall filter, switch to power Doppler, then place a spectral gate and document the absence or presence of a Doppler signal before writing 'occluded'.

Reporting lower-extremity venous reflux based on a single brief squeeze and release, or on reflux measured in a superficial or perforating vein while ignoring the deep system, and applying the same threshold to every vein segment.

Candidates learn one reflux cutoff as a number and forget that thresholds differ by vein location (deep femoral/junction vs. superficial vs. perforator) and that the augmentation maneuver technique changes the measured duration.

What to do instead

Build a one-page table of reflux thresholds by segment, and in practice perform a standardized distal augmentation with a rapid release, measuring reflux duration at each segment and timing only from the end of augmentation.

Interpreting renal artery velocities without correcting the Doppler angle or without accounting for the angle-dependent nature of velocity criteria, and accepting a technically inadequate exam as normal.

Renal arteries are small, deep, and tortuous; candidates chase a velocity number with a poor angle and then apply direct-velocity thresholds uncritically instead of relying on supporting findings.

What to do instead

In every renal practice case, document the Doppler angle used (keep it 60 degrees or less), and when the angle is unfavorable, judge by the indirect signs (tardus-parvus waveform, distal acceleration time, resistive index difference) rather than by peak velocity alone.

Setting the color scale (PRF), wall filter, or gain incorrectly for a given vessel and then diagnosing stenosis or occlusion from the resulting artifact or aliasing.

Candidates adjust knobs by habit rather than by principle, so they misread aliasing as turbulence and over-gain as flow-filled lumen, or under-gain as stenosis.

What to do instead

Practice a pre-scan checklist: set scale just above expected peak velocity, keep the Doppler angle at 60 degrees or less, adjust gain until color just begins to speckle outside the vessel, and re-check settings whenever flow appears abnormal.

Treating a screening or specialized test as a single-number result, for example reporting an ABI without confirming it with waveform analysis or without recognizing that a calcified, incompressible vessel gives a falsely elevated index.

Candidates focus on the headline number and skip the corroborating data (segmental pressures, waveforms, toe pressures) that the exam expects them to integrate.

What to do instead

For each specialized test, write the interpretation as a structured sentence that pairs the index with the waveform or pressure findings, and explicitly state when a value is unreliable (for example, ABI above 1.4 suggesting non-compressible vessels).

Answering hemodynamics or instrumentation questions by recalling a formula label rather than the physical relationship, so the answer flips when the question asks what happens to velocity when a vessel narrows or what pressure gradient means.

Memorization without a mental model of continuity and Bernoulli relationships breaks down on questions phrased as 'what happens if...' rather than 'what is the formula'.

What to do instead

Before each study session, re-derive the continuity equation and the pressure-drop relationship on paper, then work five conceptual 'what if' problems where you reason about area, velocity, and pressure rather than plugging numbers.

Classifying a carotid stenosis by memory of a single velocity cutoff while ignoring the full criteria set (plaque appearance, spectral broadening, end-diastolic velocity, post-stenotic findings) and the fact that criteria vary by laboratory.

Study guides condense stenosis into one peak systolic number, and candidates then answer questions that are deliberately testing the multi-parameter or laboratory-specific nature of the criteria.

What to do instead

Practice writing a full interpretation for each carotid case using the laboratory's stated criteria table, citing at least two parameters, and note in study notes that criteria are institution-specific and must be read from the question stem when provided.

Misidentifying which vessel is being sampled in the abdomen or pelvis, for example confusing the SMA with the celiac axis or the right renal artery with the IVC, and then applying the wrong normal values.

Candidates learn vessel names from still images but not the relational anatomy (origin level, branching pattern, relationship to aorta and pancreas) that lets them identify a vessel in real time.

What to do instead

Redraw the aortic branches with their vertebral levels and first branches from memory, and in practice identify each visceral vessel by tracing to its origin and confirming its branching pattern before taking a velocity.

Ignoring quality-assurance and artifact questions as 'not clinical', such as not recognizing that a repeat measurement discrepancy, a transducer element failure, or a machine setting drift can invalidate a study.

Candidates prioritize pathology and skip the QA domain, so they cannot connect a technical problem to the corrective action the question is asking for.

What to do instead

Study the QA domain explicitly: learn the purpose and frequency of phantom and uniformity testing, the signs of a failed element on an image, and the corrective action (repeat, recalibrate, service call) for each artifact pattern.

Exam requirements can change. Confirm current details with the exam provider.

RVT practice

200 questions. Each one lists the choices, the correct answer, and an explanation. With JavaScript, this list opens one question at a time. The checker at the top of this page saves progress in this browser when it is available.

  1. Question 1Cerebrovascular

    Which acoustic window is used to insonate the middle cerebral artery?

    1. Submandibular
    2. Transtemporal Correct answer.
    3. Transforaminal
    4. Transorbital

    Explanation

    The middle cerebral artery is reached through the thin squama of the temporal bone above the zygomatic arch, at a depth of roughly 30 to 60 mm. Each window is chosen for the vessels it can reach anatomically.

  2. Question 2Cerebrovascular

    Through the transtemporal window, the middle cerebral artery is normally sampled at a depth of approximately:

    1. 30 to 60 mm Correct answer.
    2. 5 to 15 mm
    3. 70 to 90 mm
    4. 15 to 25 mm

    Explanation

    The MCA runs laterally from the carotid terminus, so it lies 30 to 60 mm from the temporal skin surface. Depth readings help confirm vessel identity when several signals appear in the same window.

  3. Question 3Cerebrovascular

    Which Doppler angle is recommended when recording carotid velocities for stenosis grading?

    1. 60 degrees Correct answer.
    2. 30 degrees
    3. 45 degrees
    4. 90 degrees

    Explanation

    Velocities are most reproducible when the beam-to-vessel angle is kept at or below 60 degrees, and 60 degrees is the standard used in validated criteria. A 90-degree angle yields no detectable shift at all.

  4. Question 4Cerebrovascular

    Which spectral waveform is expected from a normal internal carotid artery?

    1. Low resistance with continuous forward diastolic flow Correct answer.
    2. Triphasic with early diastolic reversal
    3. Biphasic with flow reaching zero at end diastole
    4. High resistance with absent diastolic flow

    Explanation

    The internal carotid artery supplies the low-resistance cerebral circulation, so forward flow continues throughout diastole. Loss of that diastolic flow suggests distal obstruction or a hemodynamic cause.

  5. Question 5Cerebrovascular

    Tapping the superficial temporal artery produces oscillatory artifacts in which cervical vessel?

    1. Internal carotid artery
    2. Vertebral artery
    3. Common carotid artery distal to the bifurcation
    4. External carotid artery Correct answer.

    Explanation

    The superficial temporal artery is a branch of the external carotid artery, so the mechanical tap is transmitted backward into the ECA spectrum. This maneuver helps separate ECA from ICA when the two vessels run close together.

  6. Question 6Cerebrovascular

    The basilar artery is insonated through the transforaminal window at a depth of approximately:

    1. 120 to 140 mm
    2. 80 to 110 mm Correct answer.
    3. 30 to 45 mm
    4. 50 to 65 mm

    Explanation

    From the foramen magnum, the basilar artery lies deep to the clivus and is typically found at 80 to 110 mm. Velocities increase as the probe is angled toward the proximal basilar trunk.

  7. Question 7Cerebrovascular

    The pulsatility index is calculated by dividing the peak-to-trough velocity difference by which value?

    1. End-diastolic velocity
    2. Resistive index
    3. Time-averaged mean velocity Correct answer.
    4. Peak systolic velocity

    Explanation

    Gosling defined the pulsatility index as (peak systolic minus end-diastolic velocity) divided by the time-averaged mean velocity. It rises with increasing distal resistance and falls when resistance drops.

  8. Question 8Cerebrovascular

    In a normal orbit, Doppler flow in the ophthalmic artery appears:

    1. Absent during diastole
    2. Bidirectional within each cardiac cycle
    3. Antegrade and toward the probe Correct answer.
    4. Retrograde and away from the probe

    Explanation

    The ophthalmic artery carries blood from the internal carotid toward the orbit, so its normal direction is toward the transorbital probe. Flow direction here is a useful marker of collateral flow when the ICA is diseased.

  9. Question 9Cerebrovascular

    The Lindegaard ratio compares MCA velocity with the velocity in the ipsilateral extracranial internal carotid artery. A value greater than which threshold favors vasospasm over hyperemia?

    1. 3.0 Correct answer.
    2. 2.0
    3. 1.5
    4. 5.0

    Explanation

    A ratio above 3 supports vasospasm, because a global increase in flow raises velocities in both the intracranial and extracranial segments while spasm raises only the intracranial value. Ratios below 3 point toward hyperemia.

  10. Question 10Cerebrovascular

    In aneurysmal subarachnoid hemorrhage, TCD surveillance for vasospasm is most productive during which interval?

    1. The first 6 hours after the hemorrhage
    2. Weeks 4 to 6 after the hemorrhage
    3. Beginning 3 months after the hemorrhage
    4. Days 3 to 14 after the hemorrhage Correct answer.

    Explanation

    Vasospasm usually begins after the third day, peaks around days 7 to 10, and resolves by about day 14. Surveillance protocols therefore concentrate daily or alternate-day studies in that window.

  11. Question 11Cerebrovascular

    Reversed flow in the ophthalmic artery most suggests:

    1. Aneurysm arising at the basilar artery apex
    2. A distal vertebral artery stenosis at the V4 segment
    3. Ipsilateral severe ICA stenosis or occlusion Correct answer.
    4. Diffuse elevation of intracranial pressure alone

    Explanation

    When the internal carotid cannot supply the orbit, blood enters through external carotid collaterals and flows backward through the ophthalmic artery. The direction change is therefore a sign of hemodynamically significant ipsilateral ICA disease.

  12. Question 12Cerebrovascular

    In subclavian steal syndrome, the ipsilateral vertebral artery typically shows:

    1. High-velocity forward flow with marked spectral broadening
    2. Retrograde flow, at first only in diastole Correct answer.
    3. Absent diastolic flow with a sharp systolic peak
    4. Normal continuous forward flow throughout

    Explanation

    A proximal subclavian stenosis lowers pressure distal to the lesion, so the vertebral artery acts as a conduit feeding the arm. Reversal begins in diastole and becomes continuous as the pressure gradient grows.

  13. Question 13Cerebrovascular

    A young woman has alternating narrowing and dilatation along the mid-cervical internal carotid artery, with sparing of the bulb. This appearance is characteristic of:

    1. Atherosclerotic plaque at the bulb
    2. Radiation-induced arteriopathy
    3. Fibromuscular dysplasia Correct answer.
    4. Carotid body tumor

    Explanation

    Fibromuscular dysplasia produces the classic string-of-beads pattern in the mid and distal cervical ICA while the bulb is spared. Atherosclerosis, by contrast, favors the proximal bulb and bifurcation.

  14. Question 14Cerebrovascular

    A tapered flame-shaped narrowing that begins about 2 cm above the carotid bulb, accompanied by an echogenic intramural hematoma, is most consistent with:

    1. Takayasu arteritis
    2. Atherosclerotic plaque rupture
    3. Vasospasm of the cervical ICA
    4. Internal carotid artery dissection Correct answer.

    Explanation

    Dissection typically begins several centimeters above the bulb and tapers smoothly as blood tracks within the vessel wall. The intramural hematoma is the defining finding and helps separate dissection from atheroma.

  15. Question 15Cerebrovascular

    Which Doppler measurement serves as the primary criterion for grading internal carotid artery stenosis?

    1. Internal carotid end-diastolic velocity alone
    2. Common carotid end-diastolic velocity
    3. Internal carotid peak systolic velocity Correct answer.
    4. Ratio of ICA to ECA peak systolic velocity

    Explanation

    Internal carotid peak systolic velocity is the primary parameter in validated duplex criteria, with end-diastolic velocity and ICA-to-common-carotid ratio as supporting measures. Secondary parameters are used to confirm or adjust the category.

  16. Question 16Cerebrovascular

    Microembolic signals detected by transcranial Doppler are best described as:

    1. Continuous broad-band signal replacing the normal waveform
    2. Low-intensity bidirectional signals lasting several seconds
    3. Brief, unidirectional high-intensity signals occurring at random Correct answer.
    4. Symmetric signals appearing only at end diastole

    Explanation

    Embolic signals are short, unidirectional, intensity-increasing events that appear at random within an otherwise normal spectrum. They are distinguished from artifact, which is usually bidirectional and extends beyond the spectral window.

  17. Question 17Cerebrovascular

    In a patient with occlusion of one internal carotid artery, flow in the ipsilateral anterior cerebral artery is reversed. This indicates collateral supply through the:

    1. Anterior communicating artery Correct answer.
    2. Posterior communicating artery
    3. Ophthalmic artery
    4. Ipsilateral external carotid branches

    Explanation

    Blood crosses from the patent side through the anterior communicating artery and fills the ipsilateral ACA in retrograde fashion. Reversal of ACA direction is therefore a sign of effective anterior collateralization.

  18. Question 18Cerebrovascular

    Markedly raised intracranial pressure produces which middle cerebral artery waveform pattern on TCD?

    1. Continuous venous-like flow without pulsatility
    2. High resistance with reduced or reversed diastolic flow Correct answer.
    3. Low resistance with increased diastolic velocity
    4. Normal waveform with a doubled systolic velocity

    Explanation

    As intracranial pressure rises, the diastolic pressure gradient across the cerebral bed falls, so diastolic flow diminishes and may reverse. The waveform becomes progressively more pulsatile as perfusion pressure drops.

  19. Question 19Cerebrovascular

    Carotid endarterectomy confers the greatest absolute benefit in which group?

    1. Symptomatic patients with 30% stenosis
    2. Asymptomatic patients with 80% stenosis
    3. Symptomatic patients with 70% or greater stenosis Correct answer.
    4. Asymptomatic patients with 50% stenosis

    Explanation

    Benefit rises steeply with stenosis severity in symptomatic patients, and the greatest reduction in stroke risk is seen at 70% or greater narrowing. Below 50% in symptomatic patients, medical therapy is preferred.

  20. Question 20Cerebrovascular

    The distal cervical internal carotid artery near the skull base is best approached with which window?

    1. Transtemporal
    2. Transforaminal
    3. Transorbital
    4. Submandibular Correct answer.

    Explanation

    Angling the probe medially under the mandible brings the high cervical ICA into view before it enters the carotid canal. The standard anterior cervical approach often cannot reach this segment because of the mandible.

  21. Question 21Cerebrovascular

    A highly vascular mass splays the internal and external carotid arteries apart at the bifurcation. The most likely diagnosis is:

    1. Carotid body tumor Correct answer.
    2. Carotid dissection with pseudoaneurysm
    3. Atherosclerotic plaque with neovascularization
    4. Subclavian steal syndrome

    Explanation

    Carotid body tumors are paragangliomas that arise in the bifurcation and widen the angle between the ICA and ECA. They are hypervascular with low-resistance arterial flow and are not atherosclerotic in origin.

  22. Question 22Cerebrovascular

    Duplex shows a markedly narrowed ICA lumen containing a thin string of flow, with a peak systolic velocity of only 85 cm/s. The best interpretation is:

    1. Complete occlusion of the ICA
    2. Near-occlusion with reduced flow beyond a critical stenosis Correct answer.
    3. Normal variant caused by low cardiac output
    4. No hemodynamically significant stenosis

    Explanation

    At near-occlusion, flow is so limited that velocities fall despite extreme anatomic narrowing, so a low velocity must not be read as mild disease. The visible string of flow confirms the artery is patent but critically narrowed.

  23. Question 23Cerebrovascular

    Which TCD finding supports a diagnosis of brain death?

    1. Small systolic spikes and absent diastolic flow intracranially Correct answer.
    2. High MCA velocity with a Lindegaard ratio above 3
    3. Increased diastolic flow with an overall low pulsatility index
    4. A normal waveform with mean velocity reduced by half

    Explanation

    When cerebral perfusion pressure falls below intracranial pressure, flow stops in diastole and only brief systolic spikes remain. The pattern must be bilateral and found in all intracranial vessels to be supportive.

  24. Question 24Cerebrovascular

    A breath-holding index below 0.69 %/s during TCD testing indicates:

    1. A positive right-to-left shunt study
    2. Normal cerebrovascular reserve
    3. Impaired cerebrovascular reserve Correct answer.
    4. Artifact from a poor acoustic window

    Explanation

    The index expresses the percentage rise in MCA mean velocity per second of breath holding, so a blunted response means the vessels cannot dilate further. This implies the cerebral circulation is already maximally vasodilated.

  25. Question 25Cerebrovascular

    During a contrast TCD study for right-to-left shunt, a positive result is indicated by:

    1. Microembolic signals in the MCA shortly after injection Correct answer.
    2. A sudden fall in MCA velocity to zero
    3. Loss of the diastolic component of the MCA waveform
    4. Reversal of flow in the basilar artery

    Explanation

    Agitated saline crosses a shunt and appears in the cerebral circulation as countable high-intensity signals. The number and timing of signals relative to the injection grade the shunt magnitude.

  26. Question 26Cerebrovascular

    On transcranial Doppler, the mean flow velocity in a normal adult middle cerebral artery most typically falls within which range?

    1. 20 to 30 cm/s
    2. 180 to 200 cm/s
    3. 55 to 65 cm/s Correct answer.
    4. 120 to 140 cm/s

    Explanation

    Normal middle cerebral artery mean velocities in adults are roughly 55 to 65 cm/s, with an accepted normal band of about 40 to 80 cm/s. Values below this band raise concern for a proximal flow-limiting lesion or a poor acoustic window, while values well above it raise concern for stenosis, vasospasm, or hyperemia. The middle cerebral artery is the most reproducible intracranial vessel for serial velocity comparison.

  27. Question 27Cerebrovascular

    Which spectral Doppler feature distinguishes the internal carotid artery from the external carotid artery at the bifurcation?

    1. Continuous forward flow throughout diastole Correct answer.
    2. A sharp upstroke with no diastolic flow at all
    3. Prominent early diastolic flow reversal
    4. A monophasic waveform resembling venous flow

    Explanation

    The internal carotid artery perfuses the low-resistance cerebral circulation, so it shows broad spectral broadening and continuous forward flow throughout diastole. The external carotid artery supplies the higher-resistance facial and scalp beds, giving a sharper upstroke, a narrower spectral window, and low or briefly reversed diastolic flow. Temporal tapping maneuvers can be used when the waveform alone is ambiguous.

  28. Question 28Cerebrovascular

    When performing carotid duplex, which angle of insonation should be maintained?

    1. Exactly 45 degrees in every case
    2. 0 degrees
    3. 60 degrees or less Correct answer.
    4. 90 degrees

    Explanation

    The Doppler shift is proportional to the cosine of the angle between the beam and the direction of flow. At 60 degrees the cosine is 0.5 and small angle errors cause only modest velocity error, but as the angle approaches 90 degrees the cosine falls steeply and the measured shift collapses toward zero. Standard practice therefore keeps the angle at 60 degrees or less and steers the cursor to parallel the vessel wall.

  29. Question 29Cerebrovascular

    Where is the transtemporal acoustic window for transcranial Doppler located?

    1. At the angle of the mandible
    2. Just above the orbital rim
    3. Above the zygomatic arch, anterior to the ear Correct answer.
    4. At the occipital midline just below the inion

    Explanation

    The transtemporal window uses the thin squamosal portion of the temporal bone above the zygomatic arch and slightly anterior to the ear. It provides access to the terminal internal carotid artery, middle cerebral artery, anterior cerebral artery, posterior cerebral artery, and the anterior communicating region. Failure to find the window may reflect thickening of the temporal bone rather than absence of flow.

  30. Question 30Cerebrovascular

    An older adult with new temporal headache and jaw claudication undergoes a superficial temporal artery ultrasound. A hypoechoic wall thickening encircles the vessel and the lumen compresses when pressure is applied with the probe. This finding is termed:

    1. The double-line sign
    2. The yin-yang sign
    3. The sparkle sign
    4. The halo sign Correct answer.

    Explanation

    Concentric hypoechoic mural thickening of a temporal artery branch is the halo sign of giant cell arteritis, representing mural edema. A compressible lumen under probe pressure (the compression sign) adds specificity for active arteritis. Prompt recognition matters because treatment is started on clinical and sonographic grounds before biopsy results return.

  31. Question 31Cerebrovascular

    In the STOP protocol for children with sickle cell disease, which transcranial Doppler measurement identifies the group at highest stroke risk?

    1. Resistive index above 0.8
    2. Time-averaged mean of 200 cm/s or more Correct answer.
    3. Mean velocity below 70 cm/s
    4. Peak systolic velocity of 120 cm/s

    Explanation

    The STOP trial stratified stroke risk by the time-averaged mean of the maximum velocity in the middle cerebral artery or terminal internal carotid artery. A value of 200 cm/s or greater defined the high-risk group in whom regular transfusion produced a large reduction in stroke. Intermediate values of roughly 170 to 199 cm/s were designated conditional and followed with repeat studies.

  32. Question 32Cerebrovascular

    Two patients have carotid stenoses of equal angiographic severity. One plaque is uniformly echolucent and the other densely calcified with strong acoustic shadowing. Which statement best reflects the recognized clinical implication?

    1. The calcified plaque is more prone to rupture because it is brittle
    2. Plaque echogenicity has no relationship to symptoms
    3. Both plaques carry identical risk because the degree of stenosis is the same
    4. The echolucent plaque has a higher rupture risk Correct answer.

    Explanation

    Plaque composition influences risk independently of stenosis severity. Echolucent plaques are typically lipid-rich or contain hemorrhage with a thin fibrous cap, features associated with instability, rupture, and ipsilateral ischemic symptoms. Densely calcified plaques are generally more stable and less frequently the embolic source.

  33. Question 33Cerebrovascular

    During transcranial Doppler monitoring after subarachnoid hemorrhage, which mean middle cerebral artery velocity is conventionally used as the threshold suggesting vasospasm?

    1. 70 cm/s
    2. 90 cm/s
    3. 120 cm/s Correct answer.
    4. 40 cm/s

    Explanation

    A mean middle cerebral artery velocity of about 120 cm/s or greater is the conventional transcranial Doppler threshold for suspected vasospasm. Velocities above this value prompt closer correlation with the clinical examination and with the extracranial internal carotid velocity to separate spasm from hyperemia. Serial trends are more informative than a single measurement.

  34. Question 34Cerebrovascular

    Duplex of the common carotid artery shows a delayed systolic upstroke, a rounded systolic peak, and reduced overall amplitude. A severe stenosis is present in the ipsilateral brachiocephalic trunk. This waveform pattern is called:

    1. Tardus parvus Correct answer.
    2. Spectral broadening
    3. Pulsus bisferiens
    4. Dicrotic notch reversal

    Explanation

    A dampened, delayed systolic upstroke with loss of the sharp peak downstream of a hemodynamically significant proximal obstruction is the tardus parvus waveform. It reflects reduced pulse pressure reaching the distal vessel and should direct attention to the more proximal inflow vessels rather than to the segment being sampled. Recognizing it prevents misattributing the abnormality to the vessel under the probe.

  35. Question 35Cerebrovascular

    A patient has a documented occlusion of the left internal carotid artery. Duplex of the right internal carotid artery shows a peak systolic velocity of 130 cm/s with no visible plaque and a normal-appearing lumen. The most likely explanation is:

    1. Reduced cardiac output producing turbulence
    2. A false elevation caused by an incorrect Doppler angle
    3. Compensatory increased flow through the patent artery Correct answer.
    4. Interval development of a new atherosclerotic stenosis

    Explanation

    When one internal carotid artery occludes, the contralateral artery must supply both hemispheres and its volume flow rises. Velocities therefore increase without any atherosclerotic narrowing, and the spectrum keeps a normal low-resistance shape without post-stenotic disturbance. Velocity criteria for stenosis must always be interpreted with the contralateral status in mind.

  36. Question 36Cerebrovascular

    Through the temporal window, a sample at 62 mm depth shows flow directed toward the transducer, and a sample at 75 mm with slight angulation shows flow directed away. These signals most likely represent:

    1. Middle cerebral artery superficially and anterior cerebral artery deeper Correct answer.
    2. Posterior cerebral artery superficially and middle cerebral artery deeper
    3. Anterior cerebral artery at both depths
    4. Middle cerebral artery at both depths

    Explanation

    At the transtemporal window the middle cerebral artery is normally found at roughly 40 to 65 mm with flow toward the probe. Deeper and slightly anterior, the A1 segment of the anterior cerebral artery shows flow away from the probe. Depth plus direction together allow reliable vessel identification.

  37. Question 37Cerebrovascular

    A vertebral artery measures 2 mm in diameter along its entire cervical course while the contralateral vertebral artery measures 4.5 mm; velocities in the small vessel are normal to mildly elevated. This appearance is most consistent with:

    1. Subclavian steal
    2. Fibromuscular dysplasia
    3. Vertebral artery dissection
    4. Congenital vertebral artery hypoplasia Correct answer.

    Explanation

    A uniformly small-caliber vertebral artery with a smooth wall and preserved, sometimes mildly elevated, velocities throughout its course represents congenital hypoplasia. It is a common normal variant and is distinguished from acquired disease by its long-segment uniformity and absence of mural abnormality. The contralateral vessel is typically dominant and supplies most of the basilar flow.

  38. Question 38Cerebrovascular

    Transcranial Doppler insonation of a feeding artery to a cerebral arteriovenous malformation typically shows:

    1. Absent diastolic flow
    2. High velocity with low pulsatility Correct answer.
    3. Reciprocal systolic flow reversal
    4. Low velocity with high pulsatility

    Explanation

    Arteriovenous malformations shunt blood directly from artery to vein, bypassing the high-resistance capillary bed. Feeding arteries therefore carry high-volume, low-resistance flow with elevated velocities and continuous high diastolic flow, so pulsatility is reduced. These features help distinguish a feeding artery from normal parenchymal vessels.

  39. Question 39Cerebrovascular

    After microbubble contrast administration, fine mobile echoes are seen moving within the body of a soft carotid plaque. This finding most likely represents:

    1. Intraplaque hemorrhage with clot retraction
    2. Artifact from adjacent calcific shadowing
    3. Neovascularization within the plaque Correct answer.
    4. Surface ulceration of the plaque

    Explanation

    Contrast ultrasound can depict microbubbles within the fragile capillary network that grows into a thickened plaque. Visible intraplaque contrast enhancement indicates neovascularization, a marker of plaque vulnerability and prior or future ischemic symptoms. The enhancement occurs within the plaque substance rather than at its surface.

  40. Question 40Cerebrovascular

    A young patient with recurrent transient ischemic attacks has stenotic terminal internal carotid arteries with a hazy basal network of small collateral vessels. Corresponding transcranial Doppler findings would most likely include:

    1. Reversed flow in the basilar artery
    2. Uniformly increased velocities in all intracranial arteries
    3. Absent flow signals in all intracranial vessels
    4. High-velocity low-resistance flow in basal collateral vessels Correct answer.

    Explanation

    In moyamoya disease, progressive stenosis of the distal internal carotid arteries is accompanied by lenticulostriate and thalamoperforating collaterals that supply the ischemic territory. These small collateral channels carry increased volume flow at low resistance, so transcranial Doppler shows high-velocity, low-pulsatility signals near the skull base alongside stenotic velocities in the terminal carotid. This combination of proximal stenosis with collateral flow is the characteristic pattern.

  41. Question 41Peripheral Arterial

    A normal resting ankle-brachial index in a healthy adult falls within which range?

    1. 0.50 to 0.89
    2. 1.50 to 2.00
    3. Below 0.50
    4. 0.90 to 1.40 Correct answer.

    Explanation

    Normal ABI is generally 0.90 to 1.40. Lower values suggest obstructive peripheral arterial disease, while markedly elevated values suggest noncompressible calcified arteries.

  42. Question 42Peripheral Arterial

    Which spectral Doppler feature is characteristic of a normal triphasic lower extremity arterial waveform?

    1. Absent diastolic flow and rounded upstroke
    2. Broad spectral broadening in systole
    3. Continuous forward flow throughout diastole
    4. Sharp upstroke with early diastolic reversal Correct answer.

    Explanation

    A normal triphasic waveform has a steep systolic acceleration, a narrow spectral band, early diastolic flow reversal, and a small late diastolic forward component. Loss of these features suggests proximal obstruction or distal vasodilation.

  43. Question 43Peripheral Arterial

    Which cuff size is most appropriate for measuring systolic pressure at the ankle?

    1. A forearm cuff wrapped around the calf
    2. A digital cuff placed on the great toe
    3. A cuff matched to ankle circumference Correct answer.
    4. A thigh cuff placed above the knee

    Explanation

    Cuff width should be matched to limb circumference to avoid falsely high or low pressure readings. An ankle cuff selected for the patient's ankle size gives the most reliable segmental pressure.

  44. Question 44Peripheral Arterial

    The dorsalis pedis artery is the distal continuation of which artery?

    1. Posterior tibial artery
    2. Anterior tibial artery Correct answer.
    3. Popliteal artery
    4. Peroneal artery

    Explanation

    The anterior tibial artery crosses the interosseous membrane and continues onto the dorsum of the foot as the dorsalis pedis artery. The posterior tibial artery continues as the plantar arteries.

  45. Question 45Peripheral Arterial

    An ankle-brachial index of 0.80 in a symptomatic patient most likely indicates which condition?

    1. Normal arterial function
    2. Acute compartment syndrome
    3. Peripheral arterial disease Correct answer.
    4. Venous insufficiency

    Explanation

    An ABI below 0.90 is abnormal and supports a diagnosis of peripheral arterial disease. Values from 0.50 to 0.89 generally correspond to mild to moderate disease.

  46. Question 46Peripheral Arterial

    Why is the toe-brachial index often used when ankle pressures may be falsely elevated?

    1. Toe vessels are more prone to vasospasm
    2. Toe cuffs are larger than ankle cuffs
    3. Digital arteries are less often calcified Correct answer.
    4. Toe pressures directly measure iliac stenosis

    Explanation

    Medial calcinosis can make tibial arteries noncompressible and falsely elevate ankle pressures. Digital arteries are less commonly affected by calcification, so toe pressures may better reflect true perfusion.

  47. Question 47Peripheral Arterial

    Which artery is commonly used as the proximal reference for lower extremity Doppler waveforms at the groin?

    1. Profunda femoris artery
    2. Superficial femoral artery
    3. Popliteal artery
    4. Common femoral artery Correct answer.

    Explanation

    The common femoral artery is the standard proximal sampling site because it reflects inflow from the iliac system and is readily accessible at the groin. Waveform changes there help distinguish inflow from outflow disease.

  48. Question 48Peripheral Arterial

    A monophasic Doppler waveform in the dorsalis pedis artery most suggests which finding?

    1. Venous obstruction
    2. Compensatory high flow
    3. A proximal stenosis Correct answer.
    4. Normal resting flow

    Explanation

    A monophasic waveform with loss of the normal multiphasic pattern indicates damping from a proximal stenosis or occlusion. The more proximal the obstruction, the more blunted the distal waveform becomes.

  49. Question 49Peripheral Arterial

    An older patient with diabetes has an ankle-brachial index of 1.55 and nonhealing foot ulcers. What is the most likely explanation?

    1. Exercise-induced vasodilation
    2. Severe iliac stenosis limiting inflow
    3. Medial calcinosis of tibial arteries Correct answer.
    4. Improper cuff placement on the thigh

    Explanation

    An ABI above 1.40 often reflects medial calcinosis, especially in patients with diabetes or renal disease, making tibial arteries noncompressible. In this setting, toe pressures or waveform analysis are more reliable.

  50. Question 50Peripheral Arterial

    During lower extremity segmental pressure testing, a systolic pressure drop of how much between adjacent cuffs is considered hemodynamically significant?

    1. 5 mm Hg
    2. 20 mm Hg Correct answer.
    3. 60 mm Hg
    4. 40 mm Hg

    Explanation

    A drop of 20 mm Hg or more between adjacent segmental cuffs suggests a significant stenosis or occlusion in the intervening arterial segment. Smaller differences may fall within normal measurement variability.

  51. Question 51Peripheral Arterial

    A focal increase in peak systolic velocity with a velocity ratio of 2.5 across a femoral artery stenosis is most consistent with which degree of stenosis?

    1. Total occlusion
    2. 50% or greater Correct answer.
    3. 20% to 49%
    4. Less than 20%

    Explanation

    A velocity ratio of at least 2.0 at the stenosis compared with the proximal normal artery supports a diameter reduction of 50% or greater. Higher ratios generally indicate tighter stenosis.

  52. Question 52Peripheral Arterial

    A tardus-parvus waveform is detected distal to a severe stenosis. Which finding characterizes it?

    1. Early diastolic flow reversal
    2. Sharp systolic upstroke with a high peak velocity
    3. Slow systolic upstroke with low peak velocity Correct answer.
    4. High-resistance triphasic flow

    Explanation

    Tardus-parvus means a delayed, rounded systolic upstroke with reduced peak velocity. It reflects downstream damping of flow from a proximal stenosis.

  53. Question 53Peripheral Arterial

    Exercise ankle-brachial index testing is most indicated when resting ABI is normal but the patient reports which symptom?

    1. Rest pain
    2. Cold sensitivity in the fingers
    3. Swelling of the calf
    4. Intermittent claudication Correct answer.

    Explanation

    Exercise testing can unmask hemodynamically significant disease when the resting ABI is normal. A postexercise ABI decrease supports a flow-limiting stenosis.

  54. Question 54Peripheral Arterial

    Which duplex finding is most characteristic of a femoral artery pseudoaneurysm after catheterization?

    1. Monophasic waveform in the distal artery
    2. Absent flow within a surrounding hematoma
    3. Yin-yang sac with a to-and-fro neck Correct answer.
    4. Continuous high-velocity flow in the native artery

    Explanation

    A pseudoaneurysm shows swirling yin-yang color within the sac and a narrow neck with to-and-fro spectral flow. This pattern represents systolic inflow into the sac and diastolic outflow back into the artery.

  55. Question 55Peripheral Arterial

    A normally functioning femoropopliteal vein bypass graft typically shows which Doppler waveform?

    1. Reversed diastolic flow only, with no forward component
    2. High-resistance triphasic flow
    3. Low-resistance flow with diastolic forward flow Correct answer.
    4. Absent diastolic flow

    Explanation

    Vein grafts to the lower limb generally show a low-resistance waveform with continuous diastolic forward flow because they supply a vasodilated distal bed. This differs from the high-resistance triphasic pattern of a normal resting native artery.

  56. Question 56Peripheral Arterial

    A young athlete has calf pain during plantarflexion. Duplex shows occlusion of the popliteal artery only with active plantarflexion. Which diagnosis is most likely?

    1. Popliteal aneurysm
    2. Buerger disease
    3. Entrapment syndrome Correct answer.
    4. Cystic adventitial disease

    Explanation

    Popliteal entrapment syndrome causes dynamic compression of the popliteal artery by abnormal gastrocnemius muscle or fibrous bands. Symptoms and arterial occlusion are reproduced with plantarflexion or dorsiflexion.

  57. Question 57Peripheral Arterial

    Reversed flow in the vertebral artery during systole on carotid duplex most suggests which condition?

    1. Vertebrobasilar thrombosis
    2. Aortic valve regurgitation
    3. Internal carotid artery stenosis
    4. Subclavian steal Correct answer.

    Explanation

    Proximal subclavian stenosis or occlusion lowers pressure distal to the lesion, causing blood to flow retrogradely down the vertebral artery to supply the arm. This is the classic subclavian steal pattern.

  58. Question 58Peripheral Arterial

    A patient with episodic finger blanching in the cold has normal proximal upper extremity pulses. Which diagnosis best fits?

    1. Raynaud phenomenon Correct answer.
    2. Subclavian artery occlusion
    3. Giant cell arteritis
    4. Thoracic outlet syndrome

    Explanation

    Raynaud phenomenon is cold- or stress-induced digital vasospasm with normal proximal pulses. Duplex may show reversible reduction of digital artery flow during an episode.

  59. Question 59Peripheral Arterial

    Pulse volume recordings in lower extremity arterial testing measure which parameter?

    1. Peak systolic velocity
    2. Transcutaneous oxygen tension
    3. Volume changes in the segment Correct answer.
    4. Arterial diameter change

    Explanation

    Pulse volume recording uses air-filled cuffs to detect volume changes in the limb segment during the cardiac cycle. The resulting waveform reflects segmental arterial perfusion.

  60. Question 60Peripheral Arterial

    Color Doppler shows a bright mosaic of alternating colors within a stenotic artery. What is the most likely cause?

    1. Doppler angle at 90 degrees
    2. Sample volume too large for the artery
    3. Wall filter set too high for the vessel
    4. Aliasing above the Nyquist limit Correct answer.

    Explanation

    Aliasing produces color wrap-around when blood velocity exceeds the Nyquist limit. Raising the pulse repetition frequency or adjusting the color scale can reduce the artifact.

  61. Question 61Peripheral Arterial

    A middle-aged man has rapidly progressive calf claudication and duplex shows a multilocular cystic structure compressing the popliteal artery. Which condition is most likely?

    1. Popliteal artery aneurysm
    2. Buerger disease
    3. Cystic adventitial disease Correct answer.
    4. Popliteal entrapment syndrome

    Explanation

    Cystic adventitial disease typically affects the popliteal artery in young to middle-aged men and causes extrinsic compression by mucin-filled cysts in the adventitia. Duplex shows cystic spaces narrowing the arterial lumen.

  62. Question 62Peripheral Arterial

    A 35-year-old heavy smoker has distal digital ischemia with normal proximal pulses and corkscrew collaterals on angiography. Which diagnosis is most likely?

    1. Atherosclerotic iliac stenosis
    2. Thoracic outlet syndrome
    3. Buerger disease Correct answer.
    4. Raynaud phenomenon

    Explanation

    Buerger disease, or thromboangiitis obliterans, affects distal small and medium vessels in young smokers and produces segmental thrombosis with corkscrew collaterals. Proximal pulses are often normal.

  63. Question 63Peripheral Arterial

    Which maneuver during upper extremity arterial duplex is used to provoke compression at the thoracic outlet?

    1. Valsalva maneuver
    2. Arm abduction and external rotation Correct answer.
    3. Wrist flexion with repeated fist clenching
    4. Trendelenburg position

    Explanation

    Thoracic outlet compression is provoked by arm abduction and external rotation, which narrows the costoclavicular space. This can reduce subclavian artery flow or produce positional waveform changes.

  64. Question 64Peripheral Arterial

    A patient returns 8 months after femoropopliteal stenting with recurrent claudication. Duplex shows a focal peak systolic velocity of 340 cm/s within the stent and a proximal velocity of 75 cm/s. What is the most likely cause?

    1. Normal stent flow
    2. In-stent restenosis Correct answer.
    3. Stent infection
    4. Distal embolization

    Explanation

    A focal velocity ratio of about 4.5 within the stent indicates a hemodynamically significant stenosis. In-stent restenosis from intimal hyperplasia is a common cause of recurrent symptoms after stenting.

  65. Question 65Peripheral Arterial

    A patient with a chronically ischemic limb has an ankle-brachial index of 0.35 and a toe pressure of 20 mm Hg. Which classification best describes this finding?

    1. Acute limb ischemia
    2. Normal limb perfusion
    3. Critical limb ischemia Correct answer.
    4. Claudication

    Explanation

    An ABI of 0.35 with a toe pressure of 20 mm Hg indicates severe limb hypoperfusion consistent with critical limb ischemia. These values are far below the thresholds associated with claudication or normal perfusion.

  66. Question 66Peripheral Arterial

    Which description best characterizes a normal resting Doppler waveform from a lower extremity artery?

    1. Systolic flow reversal followed by a rounded, low-velocity forward peak
    2. Sharp systolic peak with uninterrupted forward diastolic flow only
    3. Blunted upstroke with forward flow continuing through all of diastole
    4. Sharp systolic peak with early diastolic reversal and late forward flow Correct answer.

    Explanation

    A normal resting limb artery is a high-resistance bed, so flow is triphasic: brisk systole, a brief early diastolic reversal against the closed distal bed, and a small late diastolic forward wave. Loss of the reversal or a blunted upstroke signals proximal disease or downstream vasodilation.

  67. Question 67Peripheral Arterial

    During ankle-brachial index measurement, a cuff that is too narrow for the limb will have which effect on the recorded systolic pressure?

    1. It falsely elevates the measured systolic pressure Correct answer.
    2. It falsely lowers the measured systolic pressure
    3. It alters only the diastolic component of the signal
    4. It has no measurable effect on the systolic pressure

    Explanation

    Cuff width must be roughly 20 percent greater than the limb diameter so that inflation pressure is transmitted to the artery. An undersized cuff requires more pressure to occlude the vessel, so the reading is falsely high; an oversized cuff produces falsely low readings.

  68. Question 68Peripheral Arterial

    Which ankle-brachial index range is considered normal in a healthy adult?

    1. 1.00 to 1.40 Correct answer.
    2. 0.75 to 0.95
    3. 0.50 to 0.70
    4. Greater than 1.50

    Explanation

    In healthy adults ankle systolic pressure slightly exceeds brachial pressure, giving a normal index of approximately 1.00 to 1.40. Values of 0.90 or below indicate obstruction, while values above 1.40 raise concern for noncompressible, calcified arteries.

  69. Question 69Peripheral Arterial

    How should the patient be positioned for lower extremity segmental pressure measurements?

    1. Seated with the legs dependent over the table edge
    2. Standing with full weight borne on the tested leg
    3. Reverse Trendelenburg at a 30-degree incline
    4. Supine with the limbs at approximately heart level Correct answer.

    Explanation

    Segmental pressures are obtained with the patient supine so that hydrostatic effects are minimized and all cuffs sit at roughly the same level as the heart. Testing in a dependent or standing position introduces column pressure that falsely raises readings in the lower limbs.

  70. Question 70Peripheral Arterial

    An interarm systolic pressure difference of at least how much suggests a hemodynamically significant subclavian artery stenosis?

    1. 70 mm Hg
    2. 15 mm Hg Correct answer.
    3. 50 mm Hg
    4. 5 mm Hg

    Explanation

    A bilateral arm pressure difference of roughly 15 to 20 mm Hg or greater is the accepted threshold for a hemodynamically significant subclavian or innominate stenosis. Smaller differences fall within normal variation, and larger differences are unusual but still simply indicate significance rather than a separate category.

  71. Question 71Peripheral Arterial

    Duplex identifies a 2.2 cm popliteal artery aneurysm. Which additional evaluation is most important?

    1. Duplex of the contralateral popliteal artery and the abdominal aorta Correct answer.
    2. Carotid duplex to exclude a tandem stenosis
    3. Renal artery duplex to exclude fibromuscular dysplasia
    4. Venous reflux study of the ipsilateral great saphenous vein

    Explanation

    Popliteal aneurysms are bilateral in roughly half of affected patients and are associated with abdominal aortic aneurysm in a substantial minority, so the opposite popliteal artery and the aorta must be imaged. Missing a synchronous aneurysm leaves the patient at risk for thrombosis or rupture at a second site.

  72. Question 72Peripheral Arterial

    After endovascular aneurysm repair, duplex shows flow within the aneurysm sac fed by a patent lumbar artery, with no flow at the attachment sites. How is this classified?

    1. Type II endoleak Correct answer.
    2. Type IV endoleak
    3. Type I endoleak
    4. Type III endoleak

    Explanation

    Type II endoleak is retrograde filling of the sac through branch vessels such as lumbar or inferior mesenteric arteries and is the most common type. Type I arises at the proximal or distal attachment zone, type III from a fabric defect or component separation, and type IV from graft porosity.

  73. Question 73Peripheral Arterial

    A femoropopliteal vein graft is widely patent with no focal velocity step-up, but mid-graft peak systolic velocity is 30 cm/s. What does this most likely represent?

    1. A low-flow state predisposing to thrombosis Correct answer.
    2. Normal graft hemodynamics
    3. A focal stenosis at the proximal anastomosis
    4. An arterialized venous valve within the graft

    Explanation

    A normally functioning infrainguinal vein graft generally carries a peak systolic velocity above about 45 cm/s throughout its length. Uniformly low velocities without a focal jet indicate a low-flow conduit, which carries an increased risk of graft failure and warrants closer surveillance.

  74. Question 74Peripheral Arterial

    A man who habitually uses the heel of his hand as a hammer reports hand pain, cold intolerance, and fingertip ulceration. Duplex shows a small aneurysm of an artery in the palm. Which artery is most likely involved?

    1. Princeps pollicis artery
    2. Superficial palmar branch of the ulnar artery Correct answer.
    3. Anterior interosseous artery
    4. Radial artery at the anatomic snuffbox

    Explanation

    Repetitive blunt trauma to the hypothenar eminence injures the superficial palmar branch of the ulnar artery as it passes through Guyon canal, producing aneurysmal degeneration or occlusion with distal embolization. This constellation is termed hypothenar hammer syndrome.

  75. Question 75Peripheral Arterial

    When calculating the ankle-brachial index for a given leg, which ankle pressure is used?

    1. The lower of the dorsalis pedis and posterior tibial pressures
    2. The mean of the dorsalis pedis and posterior tibial pressures
    3. The dorsalis pedis pressure alone
    4. The higher of the dorsalis pedis and posterior tibial pressures Correct answer.

    Explanation

    Standard technique uses the higher of the two ankle pressures in each leg and the higher of the two brachial pressures, since the higher value best reflects the true perfusion pressure to that limb. Using the lower value overestimates disease severity and reduces reproducibility.

  76. Question 76Peripheral Arterial

    With an occluded superficial femoral artery, perfusion of the calf depends most on which collateral route?

    1. Middle genicular artery arising from the popliteal artery
    2. Branches of the profunda femoris joining the genicular arteries Correct answer.
    3. Inferior epigastric artery
    4. Superficial circumflex iliac artery

    Explanation

    The profunda femoris gives off perforating branches that anastomose with the genicular arteries around the knee, providing the main alternate route to the popliteal and tibial vessels when the superficial femoral artery is occluded. This is why a patent profunda is a key determinant of limb viability in femoropopliteal disease.

  77. Question 77Peripheral Arterial

    Spectral broadening that fills the spectral window at the site of a focal stenosis reflects which flow characteristic?

    1. Venous contamination of the arterial signal
    2. Turbulent flow with a wide range of velocities Correct answer.
    3. Absence of detectable flow
    4. Laminar flow with a narrow velocity range

    Explanation

    Disturbed, turbulent flow at a stenosis contains many different velocities at once, so the spectral display broadens and the clear window beneath the waveform disappears. Uniform laminar flow, in contrast, produces a narrow band of velocities and a clean spectral window.

  78. Question 78Peripheral Arterial

    Before radial artery harvest, which bedside test assesses the adequacy of collateral flow through the palmar arch?

    1. Adson maneuver
    2. Trendelenburg test
    3. Valsalva maneuver
    4. Allen test Correct answer.

    Explanation

    The Allen test occludes both the radial and ulnar arteries while the patient clenches the fist, then releases one artery and observes reperfusion of the palm. Prompt return of color confirms that the remaining artery supplies the hand adequately.

  79. Question 79Peripheral Arterial

    For reliable peak systolic velocity measurements, the Doppler angle should be maintained at or below which value?

    1. 75 degrees
    2. 20 degrees
    3. 60 degrees Correct answer.
    4. 90 degrees

    Explanation

    Velocity estimates depend on the cosine of the Doppler angle, and errors grow rapidly as the angle approaches 90 degrees. Keeping the angle at 60 degrees or less limits cosine-related error to an acceptable range.

  80. Question 80Peripheral Arterial

    Pulse volume recordings show a normal contour at the thigh cuff but a flattened tracing with a prolonged upstroke and loss of the dicrotic notch at the calf cuff. At which level is disease localized?

    1. Aortoiliac segment
    2. Between the thigh and calf cuffs Correct answer.
    3. No disease; this is a normal variant
    4. Below the ankle

    Explanation

    Each PVR tracing reflects the arterial segment between the cuff above and the cuff below it, so a normal thigh tracing with an abnormal calf tracing places the lesion between those two levels, in the superficial femoral or popliteal segment. A flattened upstroke and absent reflected wave indicate a hemodynamically significant obstruction proximal to that cuff.

  81. Question 81Peripheral Venous

    At which venous junction does the great saphenous vein terminate?

    1. Common femoral vein Correct answer.
    2. Popliteal vein
    3. Deep femoral vein
    4. External iliac vein

    Explanation

    The great saphenous vein ascends in the saphenous compartment along the medial calf and thigh and ends at the saphenofemoral junction, entering the common femoral vein about 2 to 3 cm below the inguinal ligament. This junction is a key landmark when assessing for reflux or thrombus extension.

  82. Question 82Peripheral Venous

    Which vein does the small saphenous vein typically join?

    1. Popliteal vein Correct answer.
    2. Great saphenous vein
    3. Deep femoral vein
    4. Common femoral vein

    Explanation

    The small saphenous vein runs in the posterior calf within its own fascial compartment and usually terminates at the saphenopopliteal junction into the popliteal vein. Termination level varies, so the junction should be traced rather than assumed.

  83. Question 83Peripheral Venous

    The normal Doppler waveform in the common femoral vein of a resting patient is best described as:

    1. Pulsatile at the heart rate
    2. Absent with no detectable flow
    3. Continuous and nonpulsatile
    4. Phasic with respiration Correct answer.

    Explanation

    Normal lower limb venous flow varies with respiration because intrathoracic and intra-abdominal pressure changes alter venous return. This phasicity is lost when a proximal obstruction dampens the pressure transmission.

  84. Question 84Peripheral Venous

    Which finding is the primary criterion for diagnosing an acute deep vein thrombosis on compression ultrasound?

    1. Increased color flow within the lumen
    2. Failure of the vein to compress completely Correct answer.
    3. Loss of respirophasic variation alone
    4. Dilation of the vein on grayscale alone

    Explanation

    A normal vein coapts with light transducer pressure, so persistent noncompressibility is the most reliable sign of thrombus. Loss of phasicity and color filling defects are supportive but not diagnostic on their own.

  85. Question 85Peripheral Venous

    Acute venous thrombus typically appears how on grayscale ultrasound?

    1. Brightly echogenic with posterior shadowing
    2. Hypoechoic or anechoic Correct answer.
    3. Uniformly hyperechoic and calcified
    4. Echogenic with a smooth, well-defined wall

    Explanation

    Acute thrombus is often hypoechoic or even anechoic and may be invisible on grayscale, which is why compression is essential. Echogenicity generally increases as the thrombus organizes and becomes chronic.

  86. Question 86Peripheral Venous

    What Doppler angle is recommended for venous velocity measurements?

    1. 60 degrees Correct answer.
    2. 30 degrees
    3. 90 degrees
    4. 0 degrees

    Explanation

    An angle of about 60 degrees balances adequate Doppler signal against cosine error and is the standard for venous velocity sampling. At 90 degrees the cosine term is zero and no Doppler shift is detected.

  87. Question 87Peripheral Venous

    Venous ulcers most commonly occur in which location?

    1. Plantar surface of the foot
    2. Medial supramalleolar region Correct answer.
    3. Lateral malleolus
    4. Dorsum of the foot

    Explanation

    Chronic venous hypertension produces ulceration in the gaiter area, most often just above and behind the medial malleolus. Ulcers on the toes, plantar surface, or pressure points more often have an arterial or neuropathic cause.

  88. Question 88Peripheral Venous

    In the CEAP classification, which clinical class describes an active venous ulcer?

    1. C5
    2. C2
    3. C6 Correct answer.
    4. C4

    Explanation

    The CEAP clinical classes run from C0 (no signs) through C1 telangiectasias, C2 varicose veins, C3 edema, C4 skin changes, C5 healed ulceration, and C6 active ulceration. C6 therefore denotes an open venous ulcer.

  89. Question 89Peripheral Venous

    During venous duplex, manual distal compression of the calf normally produces which response in the popliteal vein?

    1. No detectable change
    2. Sustained flow reversal
    3. Complete flow cessation
    4. A brief increase in forward flow velocity Correct answer.

    Explanation

    Distal augmentation empties the calf veins and briefly increases forward velocity in the vein being sampled. A weak or absent response suggests obstruction between the compression point and the sample volume.

  90. Question 90Peripheral Venous

    A normal Valsalva maneuver produces which finding in the common femoral vein?

    1. Cessation or brief reversal of flow Correct answer.
    2. Continuous forward flow throughout
    3. Doubling of peak velocity
    4. A marked increase in forward flow

    Explanation

    Valsalva raises intra-abdominal pressure and temporarily reduces venous return from the legs, so flow slows, stops, or reverses briefly. A sustained reversal beyond the accepted threshold indicates valvular incompetence.

  91. Question 91Peripheral Venous

    For superficial veins, reflux is diagnosed when retrograde flow lasts longer than:

    1. 0.2 seconds
    2. 0.5 seconds Correct answer.
    3. 0.35 seconds
    4. 1.0 second

    Explanation

    The commonly used thresholds are 0.5 second for superficial veins, 1.0 second for deep veins, and 0.35 second for perforating veins. Using the correct cutoff prevents overdiagnosis of clinically insignificant brief reversal.

  92. Question 92Peripheral Venous

    Reflux testing of lower extremity veins is best performed with the patient:

    1. Sitting cross-legged
    2. Prone with knee flexed
    3. Standing with weight on the contralateral leg Correct answer.
    4. Supine and flat

    Explanation

    Reflux depends on hydrostatic pressure, which is greatest in the standing position. Having the patient stand with weight on the opposite leg distends the veins being tested while avoiding compression of the examined limb.

  93. Question 93Peripheral Venous

    Which transducer is most appropriate for routine lower extremity venous duplex in an average adult?

    1. Curved array 2-5 MHz
    2. Linear array 5-12 MHz Correct answer.
    3. Continuous-wave pencil probe 8 MHz
    4. Phased array 1-3 MHz

    Explanation

    A high-frequency linear array provides the spatial resolution needed to assess compressibility and valve detail in superficial and most deep leg veins. Lower-frequency curved arrays are reserved for obese or edematous limbs where penetration is limited.

  94. Question 94Peripheral Venous

    Which set of findings is most consistent with chronic deep vein thrombosis?

    1. Normal compressibility with phasic flow
    2. Complete noncompressibility with a hypoechoic lumen
    3. Echogenic contracted vein with partial compressibility and collaterals Correct answer.
    4. Anechoic lumen and marked distension

    Explanation

    Over time thrombus becomes echogenic and contracts, the vein may shrink, and collateral channels develop; partial compressibility can persist. Acute thrombus instead distends the vein and is hypoechoic with complete noncompressibility.

  95. Question 95Peripheral Venous

    How is D-dimer best used in suspected lower extremity DVT?

    1. To localize the thrombus level
    2. To distinguish acute from chronic thrombus
    3. To confirm the diagnosis when elevated
    4. To help exclude DVT when pretest probability is low Correct answer.

    Explanation

    D-dimer has high sensitivity but low specificity, so a negative result in a low-probability patient helps exclude DVT and avoid imaging. An elevated result is nonspecific and requires ultrasound confirmation.

  96. Question 96Peripheral Venous

    Which clinical feature adds points to the Wells score for suspected DVT?

    1. Recent surgery or immobilization
    2. Bilateral ankle edema
    3. Varicose veins
    4. Superficial bruising Correct answer.

    Explanation

    The Wells model includes active cancer, paralysis or recent immobilization, bedridden status or major surgery within 12 weeks, deep venous tenderness, entire-leg swelling, calf swelling greater than 3 cm, pitting edema, collateral superficial veins, prior DVT, and an alternative diagnosis. Superficial bruising is not one of the scored items.

  97. Question 97Peripheral Venous

    A patient has a tender cord along the medial thigh with superficial thrombophlebitis of the great saphenous vein. Which assessment is most important?

    1. Evaluate the saphenofemoral junction for extension into the deep system Correct answer.
    2. Measure the arterial ankle-brachial index
    3. Repeat the examination in six months
    4. Assess the small saphenous vein only

    Explanation

    Superficial thrombophlebitis of the great saphenous vein can propagate through the saphenofemoral junction into the common femoral vein. Ultrasound should document the proximal extent of thrombus and the distance from the junction to guide management.

  98. Question 98Peripheral Venous

    Isolated thrombosis of a soleal vein is best characterized as:

    1. An arterial embolus
    2. A normal variant requiring no follow-up
    3. A calf vein thrombosis that may propagate into deep veins Correct answer.
    4. Synonymous with superficial thrombophlebitis

    Explanation

    Soleal and gastrocnemius vein thrombi are calf vein thromboses that can extend into the popliteal and deeper veins. They are therefore reported and managed as venous thromboembolism rather than dismissed as incidental.

  99. Question 99Peripheral Venous

    To optimize compression of the popliteal vein, the patient is best positioned:

    1. Standing with weight on the examined leg
    2. Sitting with the leg dependent
    3. Supine with the knee fully extended
    4. Prone or lateral decubitus with the knee slightly flexed Correct answer.

    Explanation

    Prone or lateral decubitus positioning with slight knee flexion brings the popliteal vein closer to the transducer and reduces tension in the popliteal fossa. A fully extended supine knee can make the vein deeper and harder to compress.

  100. Question 100Peripheral Venous

    The accepted threshold for reflux in a perforating vein is retrograde flow lasting longer than:

    1. 0.1 second
    2. 1.5 seconds
    3. 0.35 second Correct answer.
    4. 0.5 second

    Explanation

    Perforator reflux is defined by retrograde flow exceeding 0.35 second, a shorter cutoff than for superficial or deep veins. This reflects the short, direct course of perforators between the deep and superficial systems.

  101. Question 101Peripheral Venous

    For deep veins of the lower extremity, reflux is diagnosed when retrograde flow exceeds:

    1. 0.35 second
    2. 0.5 second
    3. 2.0 seconds
    4. 1.0 second Correct answer.

    Explanation

    Deep veins use a longer cutoff of 1.0 second because brief reversal can be normal in the larger, lower-resistance deep system. Superficial veins use 0.5 second and perforators 0.35 second.

  102. Question 102Peripheral Venous

    Continuous, nonphasic flow in the common femoral vein suggests:

    1. An arteriovenous fistula only
    2. Deep venous reflux
    3. Normal venous return
    4. Proximal venous obstruction Correct answer.

    Explanation

    When a proximal vein is obstructed, pressure changes from respiration are not transmitted distally and flow becomes continuous. An arteriovenous fistula instead produces high-velocity, pulsatile, arterialized flow.

  103. Question 103Peripheral Venous

    May-Thurner syndrome most commonly results from compression of which structure?

    1. Left femoral vein by the inguinal ligament
    2. Right common iliac vein by the left common iliac artery
    3. Left common iliac vein by the right common iliac artery Correct answer.
    4. Popliteal vein by the gastrocnemius muscle

    Explanation

    In May-Thurner syndrome the right common iliac artery crosses anterior to the left common iliac vein and compresses it against the spine. This predisposes to left-sided iliofemoral thrombosis and chronic venous obstruction.

  104. Question 104Peripheral Venous

    Phlegmasia cerulea dolens is characterized by:

    1. Mild calf discomfort with normal pulses
    2. Superficial varicosities without edema
    3. Extensive iliofemoral thrombosis with cyanosis and risk of venous gangrene Correct answer.
    4. Chronic skin pigmentation only

    Explanation

    Phlegmasia cerulea dolens is a severe form of extensive venous thrombosis, usually iliofemoral, with marked swelling, cyanosis, and compromised arterial inflow. It threatens the limb through venous gangrene and requires urgent intervention.

  105. Question 105Peripheral Venous

    A popliteal (Baker) cyst is typically located:

    1. Within the popliteal artery wall
    2. At the saphenofemoral junction
    3. In the anterior compartment of the calf
    4. Between the medial head of the gastrocnemius and the semimembranosus tendon Correct answer.

    Explanation

    A Baker cyst arises from the gastrocnemius-semimembranosus bursa and communicates with the knee joint, appearing as a fluid collection in the medial popliteal fossa. It can mimic DVT clinically and should be distinguished from true popliteal vein thrombosis.

  106. Question 106Peripheral Venous

    Which superficial vein courses anterior to the medial malleolus?

    1. Posterior tibial vein
    2. Peroneal vein
    3. Small saphenous vein
    4. Great saphenous vein Correct answer.

    Explanation

    The great saphenous vein arises from the dorsal venous arch and ascends anterior to the medial malleolus. The small saphenous vein passes posterior to the lateral malleolus, while the posterior tibial and peroneal veins are deep calf veins.

  107. Question 107Peripheral Venous

    Which finding best confirms that a vessel is a vein rather than an artery during duplex imaging?

    1. High-resistance triphasic flow
    2. Pulsatile waveform on spectral Doppler
    3. Compressibility with light probe pressure Correct answer.
    4. Thick echogenic vessel wall

    Explanation

    Veins have thin walls and collapse when light pressure is applied with the transducer. Arteries resist compression and typically show pulsatile, higher-resistance flow.

  108. Question 108Peripheral Venous

    What is the primary function of the calf muscle pump?

    1. Increases arterial inflow to the foot
    2. Regulates capillary filtration
    3. Produces venous valves
    4. Propels venous blood toward the heart Correct answer.

    Explanation

    Contraction of the calf muscles compresses the deep veins and drives blood proximally toward the heart. Competent valves prevent retrograde flow during muscle relaxation.

  109. Question 109Peripheral Venous

    The small saphenous vein normally drains into which vein?

    1. Great saphenous vein
    2. Popliteal vein Correct answer.
    3. Deep femoral vein
    4. Common femoral vein

    Explanation

    The small saphenous vein ascends in the posterior calf and joins the popliteal vein at the saphenopopliteal junction. The great saphenous vein drains into the common femoral vein.

  110. Question 110Peripheral Venous

    Which of the following is a deep vein of the calf?

    1. Cephalic vein
    2. Peroneal vein Correct answer.
    3. Great saphenous vein
    4. Small saphenous vein

    Explanation

    The deep calf veins include the paired posterior tibial, peroneal, and anterior tibial veins. The great and small saphenous veins are superficial, and the cephalic vein is in the upper extremity.

  111. Question 111Peripheral Venous

    Color Doppler shows little or no color filling in a leg vein with sluggish flow. Which adjustment is most likely to improve detection?

    1. Increase the Doppler angle to 90 degrees
    2. Increase the color box size
    3. Raise the wall filter
    4. Lower the PRF/scale Correct answer.

    Explanation

    Slow venous flow is best detected with a low PRF/scale and a low wall filter. Lowering the PRF increases sensitivity to low-velocity flow without aliasing.

  112. Question 112Peripheral Venous

    Which grayscale finding is most typical of an acute deep vein thrombosis?

    1. A hypoechoic, distended, noncompressible vein Correct answer.
    2. A small echogenic vein with surrounding collaterals
    3. A normal-caliber vein that fully compresses
    4. A calcified intraluminal mass

    Explanation

    Acute thrombus is typically hypoechoic and causes venous distension with failure to compress. Chronic thrombus tends to be echogenic and retracted with collateral formation.

  113. Question 113Peripheral Venous

    Cockett perforating veins are located in which region?

    1. Lateral thigh
    2. Dorsal foot
    3. Medial lower calf Correct answer.
    4. Popliteal fossa

    Explanation

    Cockett perforators connect the superficial great saphenous system to the deep posterior tibial veins in the medial lower calf. They are a common site of perforator incompetence.

  114. Question 114Peripheral Venous

    Normal venous valves in the lower extremity are best described as:

    1. A single flap
    2. Bicuspid leaflets Correct answer.
    3. A muscular sphincter
    4. Three leaflets

    Explanation

    Venous valves consist of two thin leaflets that coapt to prevent retrograde flow. Their competence is assessed during reflux testing.

  115. Question 115Peripheral Venous

    Which patient is at highest risk for post-thrombotic syndrome?

    1. A patient with superficial thrombophlebitis
    2. A patient with a large proximal DVT and persistent obstruction Correct answer.
    3. A patient with a normal venous duplex
    4. A patient with an isolated calf DVT treated promptly

    Explanation

    Post-thrombotic syndrome results from chronic venous hypertension after deep vein thrombosis. The risk is greatest with large proximal thrombi and persistent venous obstruction.

  116. Question 116Peripheral Venous

    Pulsatile flow in the common femoral vein is most often caused by which condition?

    1. Right heart failure Correct answer.
    2. Competent venous valves
    3. Calf muscle pump failure
    4. Deep vein thrombosis

    Explanation

    Pulsatile venous flow reflects transmitted right atrial pressure, commonly from right heart failure or tricuspid regurgitation. Deep vein thrombosis more often produces continuous or absent phasic flow.

  117. Question 117Peripheral Venous

    Which vein is the most common site for catheter-associated thrombosis?

    1. Great saphenous vein
    2. Popliteal vein
    3. Common femoral vein
    4. Subclavian vein Correct answer.

    Explanation

    Central venous catheters most often cause thrombosis in the subclavian vein and other large upper extremity central veins. Lower extremity veins are less commonly involved in catheter-related thrombosis.

  118. Question 118Peripheral Venous

    A young competitive swimmer develops sudden arm swelling, and duplex shows thrombosis of the subclavian vein. This is best termed:

    1. Phlegmasia cerulea dolens
    2. Paget-Schroetter syndrome Correct answer.
    3. May-Thurner syndrome
    4. Post-thrombotic syndrome

    Explanation

    Paget-Schroetter syndrome is effort-related thrombosis of the subclavian vein caused by repetitive arm activity. It is also called effort thrombosis.

  119. Question 119Peripheral Venous

    Venous claudication is characterized by:

    1. Cramping pain relieved by dependency
    2. Bursting pain relieved by leg elevation Correct answer.
    3. Rest pain relieved by cold
    4. Sharp pain relieved by walking

    Explanation

    Venous claudication causes a bursting, tight pain during exercise from venous hypertension and outflow obstruction. It is relieved by elevation, which lowers venous pressure.

  120. Question 120Peripheral Venous

    A technologist is asked to map the great saphenous vein before coronary artery bypass grafting. What is the primary purpose?

    1. To assess GSV caliber and patency as a conduit Correct answer.
    2. To assess valve competence at the saphenofemoral junction
    3. To measure arterial peak systolic velocity
    4. To confirm deep venous reflux

    Explanation

    Preoperative vein mapping assesses the caliber, patency, and suitability of the great saphenous vein as a bypass conduit. It also identifies segments that are too small, thrombosed, or varicose for use.

  121. Question 121Abdominal and Visceral Vascular

    An infrarenal abdominal aortic diameter of 3.2 cm is classified as:

    1. Ectatic aorta
    2. Abdominal aortic aneurysm Correct answer.
    3. Normal aorta
    4. Aortic dissection with intimal flap

    Explanation

    The threshold for an abdominal aortic aneurysm is a maximal diameter of 3.0 cm or greater. Diameters of 2.5 to 2.9 cm are considered ectatic, and less than 2.5 cm is normal. A dissection is identified by an intimal flap, not by diameter alone.

  122. Question 122Abdominal and Visceral Vascular

    Which major abdominal vessel provides the primary arterial supply to the small intestine and right colon?

    1. Superior mesenteric artery Correct answer.
    2. Celiac trunk
    3. Middle hepatic artery
    4. Inferior mesenteric artery branch

    Explanation

    The superior mesenteric artery supplies the jejunum, ileum, and right colon through its branches. The inferior mesenteric artery supplies the left colon and rectum. The celiac trunk supplies foregut derivatives, including the liver, spleen, and stomach.

  123. Question 123Abdominal and Visceral Vascular

    In a fasting patient, the superior mesenteric artery Doppler waveform typically demonstrates:

    1. High-resistance flow with low end-diastolic velocity Correct answer.
    2. Reversed flow during systole
    3. Low-resistance flow with high continuous diastolic velocity
    4. Continuous venous flow without pulsatility

    Explanation

    In the fasting state, the SMA is a high-resistance vessel with low diastolic flow. After a meal, it becomes low-resistance with increased diastolic velocity. Continuous venous flow and systolic reversal are not normal SMA fasting patterns.

  124. Question 124Abdominal and Visceral Vascular

    Normal portal vein spectral Doppler shows flow that is:

    1. Bidirectional with each heartbeat
    2. Hepatopetal, low-velocity flow Correct answer.
    3. Hepatofugal and pulsatile
    4. Continuous and arterial in shape

    Explanation

    The normal portal vein carries blood toward the liver (hepatopetal) at a low velocity, typically about 15 to 30 cm/s. It shows gentle respiratory variation rather than arterial pulsatility. Hepatofugal flow is abnormal and suggests portal hypertension or thrombosis.

  125. Question 125Abdominal and Visceral Vascular

    A tardus-parvus waveform detected downstream of a renal artery indicates:

    1. Venous congestion
    2. Acute renal vein thrombosis
    3. Normal renal perfusion
    4. Proximal arterial stenosis Correct answer.

    Explanation

    A tardus-parvus waveform has a delayed upstroke and low amplitude, reflecting reduced systolic pressure distal to a stenosis. It is an indirect sign of upstream arterial narrowing. Venous abnormalities do not produce this arterial waveform.

  126. Question 126Abdominal and Visceral Vascular

    Normal inferior vena cava spectral Doppler flow is best described as:

    1. Phasic with respiration Correct answer.
    2. High-resistance arterial flow
    3. Absent during inspiration
    4. Continuous and nonpulsatile throughout the cardiac cycle

    Explanation

    IVC flow is influenced by right atrial pressure and respiration, producing phasic changes. It is not continuous and nonpulsatile like portal venous flow. Loss of phasicity may suggest obstruction or elevated right heart pressure.

  127. Question 127Abdominal and Visceral Vascular

    The normal celiac artery Doppler waveform is characterized by:

    1. Reversed diastolic flow
    2. Low-resistance with forward diastole Correct answer.
    3. Venous-type continuous flow
    4. High-resistance flow with absent diastole

    Explanation

    The celiac artery supplies a low-resistance vascular bed, so it normally shows continuous forward diastolic flow. A high-resistance or reversed-diastolic pattern suggests stenosis or altered downstream resistance. Venous-type continuous flow lacks arterial pulsatility.

  128. Question 128Abdominal and Visceral Vascular

    The standard sonographic measurement used to size an abdominal aortic aneurysm is the:

    1. Longitudinal length from celiac to bifurcation
    2. Circumference at the level of the renal arteries
    3. Inner-to-inner diameter in the sagittal plane
    4. Maximum AP diameter on a transverse image Correct answer.

    Explanation

    Aneurysm size is reported as the maximum anteroposterior diameter measured on a transverse image. Consistency in measurement technique is important for surveillance. Length and circumference are not the standard descriptors of aneurysm size.

  129. Question 129Abdominal and Visceral Vascular

    Which duplex finding is a direct criterion for a hemodynamically significant renal artery stenosis?

    1. Tardus-parvus waveform in the aorta
    2. Renal resistive index greater than 0.70
    3. Renal vein velocity greater than 100 cm/s
    4. RAR of 3.5 or greater Correct answer.

    Explanation

    A renal-to-aortic peak systolic velocity ratio of 3.5 or greater is a direct criterion for significant renal artery stenosis. Direct criteria are measured at the stenosis itself. Resistive index and venous velocities are not direct stenosis criteria.

  130. Question 130Abdominal and Visceral Vascular

    Isolated gastric varices in a patient with a patent portal vein most suggest:

    1. Portal vein thrombosis
    2. Inferior vena cava obstruction
    3. Budd-Chiari syndrome
    4. Splenic vein thrombosis Correct answer.

    Explanation

    Splenic vein thrombosis can cause isolated left-sided portal hypertension and gastric varices while the main portal vein remains patent. Portal vein thrombosis would typically affect the main portal system. Budd-Chiari syndrome involves hepatic venous outflow obstruction.

  131. Question 131Abdominal and Visceral Vascular

    A superior mesenteric artery peak systolic velocity greater than 275 cm/s suggests:

    1. An arteriovenous fistula
    2. Normal fasting flow
    3. A postprandial hyperemic response
    4. Significant stenosis Correct answer.

    Explanation

    A peak systolic velocity above approximately 275 cm/s is a commonly used threshold for significant SMA stenosis. Postprandial hyperemia raises flow but does not usually exceed this threshold. Arteriovenous fistulas produce high venous velocities and low-resistance flow.

  132. Question 132Abdominal and Visceral Vascular

    A type II endoleak after endovascular aortic aneurysm repair is caused by:

    1. A tear in the aorta remote from the stent
    2. Retrograde lumbar or IMA branch flow Correct answer.
    3. Incomplete seal at the proximal or distal attachment site
    4. A defect in the graft fabric

    Explanation

    Type II endoleaks arise from retrograde flow through lumbar or inferior mesenteric artery branches into the aneurysm sac. Type I leaks occur at attachment sites, and type III leaks result from graft defects. Type IV leaks are due to graft porosity.

  133. Question 133Abdominal and Visceral Vascular

    After orthotopic liver transplantation, absence of hepatic artery flow on duplex is most concerning for:

    1. Budd-Chiari syndrome
    2. Hepatic artery thrombosis Correct answer.
    3. Acute cellular rejection of the graft
    4. Portal vein thrombosis

    Explanation

    Hepatic artery thrombosis is a serious early complication that can cause biliary ischemia and graft failure. Duplex shows absent or diminished hepatic artery flow. Portal vein thrombosis and Budd-Chiari affect venous outflow or inflow, not arterial signal.

  134. Question 134Abdominal and Visceral Vascular

    Which portal vein Doppler finding is most suggestive of portal hypertension?

    1. Phasic flow that varies with respiration
    2. Peak velocity of 25 cm/s with hepatopetal flow
    3. Portal vein diameter of 10 mm
    4. Hepatofugal flow with a dilated portal vein Correct answer.

    Explanation

    Hepatofugal portal flow and portal vein dilation are classic signs of portal hypertension. Normal portal flow is hepatopetal at low velocity with respiratory variation. A 10-mm portal vein is within normal limits.

  135. Question 135Abdominal and Visceral Vascular

    Budd-Chiari syndrome is characterized by obstruction of:

    1. The splenic vein at the pancreas
    2. The portal vein at the hepatic hilum
    3. The superior mesenteric artery
    4. Hepatic veins or intrahepatic IVC Correct answer.

    Explanation

    Budd-Chiari syndrome results from hepatic venous outflow obstruction, often involving the hepatic veins or intrahepatic IVC. It leads to hepatic congestion, ascites, and portal hypertension. Portal vein and splenic vein thrombosis are separate conditions.

  136. Question 136Abdominal and Visceral Vascular

    Nutcracker syndrome results from compression of which vessel?

    1. Left renal vein between SMA and aorta Correct answer.
    2. Right gonadal vein at the inguinal canal
    3. Left renal artery by the diaphragm
    4. Inferior vena cava by the liver

    Explanation

    Nutcracker syndrome is compression of the left renal vein between the SMA and the aorta. It can cause hematuria and left flank pain. Renal artery compression and IVC compression are different entities.

  137. Question 137Abdominal and Visceral Vascular

    A tardus-parvus waveform in the common femoral artery most likely indicates:

    1. Significant aortoiliac stenosis Correct answer.
    2. Normal flow distal to the inguinal ligament
    3. Common femoral vein thrombosis
    4. An arteriovenous fistula in the thigh

    Explanation

    The common femoral artery is downstream of the aortoiliac segment, so a tardus-parvus waveform suggests proximal stenosis. Venous thrombosis does not produce an arterial tardus-parvus pattern. Arteriovenous fistulas cause high-velocity, low-resistance flow.

  138. Question 138Abdominal and Visceral Vascular

    Color Doppler of a pseudoaneurysm typically demonstrates:

    1. Laminar flow with a normal triphasic waveform
    2. A high-resistance waveform without a neck
    3. A uniform echogenic thrombus without flow
    4. Yin-yang color with a to-and-fro neck Correct answer.

    Explanation

    A pseudoaneurysm shows swirling flow within the sac, creating a yin-yang color pattern, and a narrow neck with to-and-fro spectral flow. A simple hematoma would not show internal flow. Normal triphasic flow is not seen within the sac.

  139. Question 139Abdominal and Visceral Vascular

    The most common visceral artery aneurysm is located in the:

    1. Renal artery
    2. Superior mesenteric artery
    3. Hepatic artery
    4. Splenic artery Correct answer.

    Explanation

    Splenic artery aneurysms are the most common visceral artery aneurysms. They are associated with portal hypertension, pregnancy, and pancreatitis. Hepatic and SMA aneurysms are less common.

  140. Question 140Abdominal and Visceral Vascular

    Symptomatic chronic mesenteric ischemia usually requires significant stenosis or occlusion of at least:

    1. Only the celiac trunk
    2. The portal vein and superior mesenteric vein
    3. Two of the three major mesenteric vessels Correct answer.
    4. Only the inferior mesenteric artery

    Explanation

    The mesenteric circulation has extensive collaterals, so chronic symptoms usually develop when at least two of the three major vessels (celiac, SMA, IMA) are significantly diseased. Single-vessel disease is often asymptomatic. Venous disease causes a different clinical picture.

  141. Question 141Abdominal and Visceral Vascular

    A TIPS shunt with a peak velocity of 40 cm/s and no visible flow should be interpreted as:

    1. An artifact requiring no follow-up
    2. A finding consistent with portal vein thrombosis only
    3. Significant shunt stenosis or occlusion Correct answer.
    4. A normal functioning shunt

    Explanation

    Normal TIPS velocities are generally about 90 to 190 cm/s. A velocity below 50 cm/s or absent flow suggests shunt stenosis or occlusion. Such findings require further evaluation.

  142. Question 142Abdominal and Visceral Vascular

    Which Doppler finding best supports median arcuate ligament syndrome rather than atherosclerotic celiac stenosis?

    1. Poststenotic dilation of the common hepatic artery
    2. Respiratory velocity change with a hook-shaped stenosis Correct answer.
    3. Calcific plaque at the celiac origin
    4. Celiac PSV less than 200 cm/s

    Explanation

    Median arcuate ligament syndrome produces dynamic celiac compression, so velocities vary with respiration and the stenosis has a characteristic hook shape. Atherosclerotic stenosis is fixed and often shows calcific plaque. Poststenotic dilation can occur with any stenosis.

  143. Question 143Abdominal and Visceral Vascular

    Which duplex finding favors tumor thrombus over bland portal vein thrombosis?

    1. Cavernous transformation around the porta hepatis
    2. Pulsatile arterial flow within the thrombus Correct answer.
    3. Absent flow in the portal vein
    4. Anechoic thrombus within the portal vein

    Explanation

    Tumor thrombus contains neovascularity and may show pulsatile arterial flow on Doppler. Bland thrombus is avascular. Cavernous transformation indicates chronic thrombosis and does not distinguish tumor from bland thrombus.

  144. Question 144Abdominal and Visceral Vascular

    Which duplex finding confirms an abdominal aortic dissection?

    1. A peak systolic velocity of 100 cm/s in the aorta
    2. Intimal flap with two flow lumens Correct answer.
    3. A focal aortic diameter of 3.5 cm
    4. Atherosclerotic plaque at the bifurcation

    Explanation

    An intimal flap with two lumens is the defining sonographic feature of aortic dissection. Aneurysm is defined by diameter, and plaque or velocity alone does not confirm dissection. Color and spectral Doppler can show different flow in the true and false lumens.

  145. Question 145Abdominal and Visceral Vascular

    A renal transplant patient has a focal peak systolic velocity of 300 cm/s at the arterial anastomosis with a downstream tardus-parvus waveform. This most likely indicates:

    1. Transplant renal artery stenosis Correct answer.
    2. Acute renal vein thrombosis
    3. Normal post-transplant hyperemia
    4. A urinary tract obstruction

    Explanation

    A focal high velocity at the anastomosis with a downstream tardus-parvus waveform indicates transplant renal artery stenosis. The indirect downstream sign supports a hemodynamically significant lesion. Venous thrombosis and urinary obstruction do not produce this arterial pattern.

  146. Question 146Abdominal and Visceral Vascular

    Normal main portal vein flow on duplex ultrasound is:

    1. Hepatopetal with a mean velocity above 100 cm/s
    2. Bidirectional, reversing with each cardiac cycle
    3. Hepatofugal with continuous high-velocity flow
    4. Hepatopetal with a mean velocity of about 15–30 cm/s Correct answer.

    Explanation

    The normal main portal vein carries blood toward the liver (hepatopetal) at a low, fairly continuous mean velocity of roughly 15–30 cm/s. Flow varies mildly with respiration and cardiac cycle but does not reverse. Velocities well above this range or reversed direction indicate an abnormal hemodynamic state.

  147. Question 147Abdominal and Visceral Vascular

    An infrarenal abdominal aortic diameter of 3.0 cm or greater in an adult is defined as:

    1. Aortic atherosclerosis without aneurysm
    2. An aortic dissection
    3. An abdominal aortic aneurysm Correct answer.
    4. A normal aortic diameter

    Explanation

    An infrarenal aortic diameter of 3.0 cm or more, or an increase of at least 50% compared with the normal proximal segment, defines an abdominal aortic aneurysm. The measurement is taken outer wall to outer wall in the anteroposterior plane. Atherosclerosis alone does not define an aneurysm unless the diameter criterion is met.

  148. Question 148Abdominal and Visceral Vascular

    The superior mesenteric artery Doppler waveform in a fasting patient is typically:

    1. Absent end-diastolic flow
    2. Low resistance with high continuous diastolic flow
    3. Reversed flow during diastole
    4. High resistance with low diastolic flow Correct answer.

    Explanation

    In the fasting state the mesenteric bed is constricted, so the superior mesenteric artery shows a high-resistance waveform with a sharp systolic peak and low diastolic flow. After a meal the bed dilates and the waveform becomes low resistance with increased diastolic flow. This fasting-to-postprandial change is used to assess mesenteric vascular reactivity.

  149. Question 149Abdominal and Visceral Vascular

    The normal resistive index of the hepatic artery in a native liver is approximately:

    1. 0.20–0.35
    2. 1.10–1.30
    3. 0.85–1.00
    4. 0.55–0.70 Correct answer.

    Explanation

    The hepatic artery normally shows a low-resistance waveform with a resistive index of roughly 0.55–0.70, reflecting continuous diastolic flow to the liver. Values persistently above about 0.75–0.80 raise concern for parenchymal disease or rejection in a transplant. A resistive index approaching or exceeding 1.0 implies absent or reversed diastolic flow.

  150. Question 150Abdominal and Visceral Vascular

    A type I endoleak after endovascular aneurysm repair is caused by:

    1. Retrograde flow through a lumbar or inferior mesenteric artery
    2. Porosity of the graft fabric
    3. A tear in the endograft limb
    4. Incomplete seal at the proximal or distal attachment site Correct answer.

    Explanation

    Type I endoleak results from failure to obtain a seal at the proximal (Ia) or distal (Ib) attachment zone, allowing blood to enter the sac directly from the aorta or iliac artery. It is considered a high-pressure leak that usually requires reintervention. Types II, III and IV have different mechanisms and appearances.

  151. Question 151Abdominal and Visceral Vascular

    After endovascular aneurysm repair, an aneurysm sac that enlarges by more than 5 mm on follow-up imaging is most suggestive of:

    1. Measurement artifact from bowel gas
    2. Aortic dissection
    3. A persistent endoleak Correct answer.
    4. Normal sac remodeling

    Explanation

    A successfully excluded aneurysm sac should shrink or remain stable over time. Growth of more than 5 mm during follow-up indicates continued pressurization, most often from an endoleak, and warrants further evaluation. Stable or decreasing sac size is the expected favorable response.

  152. Question 152Abdominal and Visceral Vascular

    Which set of findings represents the indirect (downstream) sign of a hemodynamically significant renal artery stenosis?

    1. Peak systolic velocity of 100 cm/s at the renal artery ostium
    2. Bidirectional flow in the renal vein
    3. Renal artery resistive index of 0.60
    4. Tardus-parvus waveform with acceleration time greater than 70 ms Correct answer.

    Explanation

    A significant proximal renal artery stenosis dampens the systolic upstroke downstream, producing a tardus-parvus intrarenal waveform with prolonged acceleration time and loss of the early systolic peak. Acceleration time above about 70 ms is considered abnormal. Direct signs are based on velocity at the stenosis itself.

  153. Question 153Abdominal and Visceral Vascular

    Cavernous transformation of the portal vein on ultrasound indicates:

    1. Chronic portal vein thrombosis with collateral formation Correct answer.
    2. Portal venous gas
    3. Acute portal vein thrombosis
    4. A normal variant of portal venous anatomy

    Explanation

    Cavernous transformation consists of multiple tortuous periportal collaterals that develop over time to bypass a chronically occluded portal vein. It implies long-standing thrombosis rather than an acute event. Acute thrombosis instead shows an expanded vein containing echogenic thrombus without these collaterals.

  154. Question 154Abdominal and Visceral Vascular

    Loss of the normal triphasic pulsatility of the hepatic vein waveform in a patient with cirrhosis most likely reflects:

    1. Elevated hepatic venous pressure from parenchymal disease Correct answer.
    2. Hepatic artery stenosis
    3. Renal vein thrombosis
    4. Normal changes of aging

    Explanation

    The hepatic vein normally shows a triphasic waveform driven by right atrial pressure changes. A stiff, fibrotic liver in cirrhosis damps transmission of these pressure changes, flattening the waveform and reducing phasicity. The finding correlates with the severity of portal hypertension.

  155. Question 155Abdominal and Visceral Vascular

    The diameter threshold above which a splenic artery aneurysm is generally considered for intervention is:

    1. 1.0 cm
    2. 0.8 cm
    3. 0.5 cm
    4. 2.0 cm Correct answer.

    Explanation

    Splenic artery aneurysms are the most common visceral artery aneurysms and rupture risk rises substantially once the diameter exceeds about 2.0 cm. Pregnancy, portal hypertension and liver transplantation lower the threshold for treatment. Smaller aneurysms are usually followed with surveillance imaging.

  156. Question 156Abdominal and Visceral Vascular

    Which finding is most consistent with acute superior mesenteric vein thrombosis?

    1. A small, collapsed superior mesenteric vein with arterialized flow
    2. Reversed flow in the superior mesenteric vein with a normal caliber
    3. A compressible superior mesenteric vein with hepatopetal flow
    4. An enlarged, noncompressible superior mesenteric vein containing echogenic material with absent flow Correct answer.

    Explanation

    In acute mesenteric venous thrombosis the vein is distended and noncompressible, contains echogenic thrombus and shows absent or partial color and spectral flow. The acute swelling distinguishes it from chronic thrombosis, where the vein may be small and irregular with collaterals. Prompt recognition matters because of the risk of bowel infarction.

  157. Question 157Abdominal and Visceral Vascular

    Pulsatile portal vein flow in a patient with a normal liver is most often caused by:

    1. Portal vein thrombosis
    2. Hepatic artery thrombosis
    3. Splenic artery stenosis
    4. Right-sided heart failure with elevated right atrial pressure Correct answer.

    Explanation

    Portal venous flow normally varies only slightly with respiration and the cardiac cycle. When right atrial pressure is elevated, as in right heart failure or tricuspid regurgitation, the pressure changes are transmitted through the hepatic sinusoids and produce a pulsatile portal waveform. The liver itself may be normal in this setting.

  158. Question 158Abdominal and Visceral Vascular

    Fibromuscular dysplasia of the renal artery characteristically:

    1. Affects the mid-to-distal artery with a string-of-beads appearance in young women Correct answer.
    2. Causes uniform dilation of the entire renal artery
    3. Produces a focal eccentric plaque at the vessel origin
    4. Involves the ostium and proximal segment in older men

    Explanation

    Fibromuscular dysplasia typically spares the ostium and involves the mid-to-distal renal artery, producing the classic string-of-beads appearance from alternating stenoses and dilations. It most often affects young to middle-aged women. Atherosclerosis, by contrast, favors the ostium and proximal segment in older patients.

  159. Question 159Abdominal and Visceral Vascular

    Circumferential, homogeneous mural thickening of the abdominal aorta in a young woman with elevated inflammatory markers is most suggestive of:

    1. Retroperitoneal fibrosis
    2. Takayasu arteritis Correct answer.
    3. A mycotic aneurysm
    4. Atherosclerotic plaque

    Explanation

    Takayasu arteritis produces a circumferential, homogeneous, hypoechoic mural thickening of the aorta and its major branches in younger patients, often women, with raised inflammatory markers. The thickening is concentric and smooth rather than calcified and eccentric. It can progress to stenosis or aneurysm formation.

  160. Question 160Abdominal and Visceral Vascular

    A saccular abdominal aortic aneurysm with irregular, lobulated margins in a febrile patient with positive blood cultures most likely represents:

    1. A dissecting aneurysm
    2. A mycotic (infected) aneurysm Correct answer.
    3. An atherosclerotic fusiform aneurysm
    4. A pseudoaneurysm from prior surgery

    Explanation

    Mycotic aneurysms are infected arterial aneurysms that typically appear saccular with irregular, lobulated walls and may show periaortic fluid, gas or adjacent inflammatory change. Fever and bacteremia support the diagnosis. They tend to enlarge rapidly and carry a high risk of rupture.

  161. Question 161Instrumentation, Quality Assurance and Specialized Vascular Testing

    What is the maximum Doppler angle recommended for accurate velocity measurement with spectral Doppler?

    1. 90 degrees
    2. 45 degrees
    3. 60 degrees Correct answer.
    4. 30 degrees

    Explanation

    Velocity is calculated using the cosine of the Doppler angle, and accuracy falls as the angle approaches 90 degrees because the cosine approaches zero. An angle of 60 degrees or less keeps the cosine term large enough for acceptable accuracy, and 60 degrees is the usual maximum recommended.

  162. Question 162Instrumentation, Quality Assurance and Specialized Vascular Testing

    Aliasing on a spectral Doppler tracing is caused by which condition?

    1. Wall filter adjusted far too low
    2. Sample volume positioned too deep in tissue
    3. Doppler receiver gain set much too high
    4. Shift exceeding the Nyquist limit Correct answer.

    Explanation

    The Nyquist limit equals half the pulse repetition frequency. When the Doppler shift exceeds this limit, the displayed velocity wraps to the opposite channel, producing aliasing.

  163. Question 163Instrumentation, Quality Assurance and Specialized Vascular Testing

    Increasing the wall filter on a color Doppler image will have which effect?

    1. It suppresses low-velocity wall motion signals Correct answer.
    2. It increases detection of low-velocity flow
    3. It increases the color frame rate
    4. It eliminates aliasing in high-velocity vessels

    Explanation

    The wall filter removes low-frequency Doppler shifts, which are typically produced by moving vessel walls and cardiac pulsation. Raising the filter improves rejection of this artifact but can also suppress genuine slow flow.

  164. Question 164Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which Doppler mode does not provide depth discrimination of the detected flow?

    1. Power Doppler
    2. Color Doppler
    3. Continuous wave Doppler Correct answer.
    4. Pulsed wave Doppler

    Explanation

    Continuous wave Doppler transmits and receives continuously, so signal is collected along the entire beam path and no specific depth can be assigned. Pulsed wave techniques use range gating to select a depth.

  165. Question 165Instrumentation, Quality Assurance and Specialized Vascular Testing

    The sample volume gate in pulsed wave Doppler is used to:

    1. increase the transmitted acoustic power
    2. adjust the color map direction
    3. filter out low-frequency Doppler shifts
    4. set the depth of velocity measurement Correct answer.

    Explanation

    The sample volume gate determines the depth along the beam from which returning echoes are processed for spectral display. Moving the gate changes which vessel segment is interrogated.

  166. Question 166Instrumentation, Quality Assurance and Specialized Vascular Testing

    On a color Doppler display, variance is used to represent:

    1. flow direction relative to the transducer
    2. the depth of the sample volume
    3. the velocity scale in cm/s
    4. turbulent or disturbed flow Correct answer.

    Explanation

    Variance mapping adds a separate color (often green or yellow) when Doppler shifts within a sample are widely distributed, which indicates disturbed or turbulent flow. Direction is shown by the red-blue map, and scale and depth are separate settings.

  167. Question 167Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which control adjusts the overall brightness of the spectral Doppler waveform?

    1. Sample volume size
    2. Receiver gain Correct answer.
    3. Velocity scale
    4. Wall filter

    Explanation

    Receiver gain amplifies returning echoes and directly controls the brightness of the spectral display. Excessive gain can fill in the spectral window, while insufficient gain makes the waveform faint.

  168. Question 168Instrumentation, Quality Assurance and Specialized Vascular Testing

    A Doppler flow phantom is primarily used in quality assurance to evaluate:

    1. patient examination room shielding
    2. transducer acoustic output in milliwatts
    3. image display monitor calibration only
    4. accuracy of velocity measurements Correct answer.

    Explanation

    A flow phantom produces a known flow velocity or Doppler shift, allowing the system's velocity measurement accuracy to be verified. It may also be used for sensitivity and resolution testing, but velocity accuracy is a core purpose.

  169. Question 169Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which adjustment would most directly reduce aliasing in a spectral Doppler tracing of a high-velocity artery?

    1. Decrease the Doppler angle to 90 degrees
    2. Increase the wall filter
    3. Increase the receiver gain
    4. Increase the pulse repetition frequency Correct answer.

    Explanation

    Aliasing occurs when the Doppler shift exceeds the Nyquist limit, which is half the pulse repetition frequency. Raising the PRF raises the Nyquist limit and expands the displayable velocity range.

  170. Question 170Instrumentation, Quality Assurance and Specialized Vascular Testing

    Increasing the transmitted Doppler frequency from 3 MHz to 7 MHz will have which effect?

    1. No change in Doppler shift but improved range resolution
    2. A higher Doppler shift and reduced penetration Correct answer.
    3. A higher Doppler shift and greater penetration
    4. A lower Doppler shift and greater penetration

    Explanation

    Doppler shift is directly proportional to the transmitted frequency, so a higher frequency produces a larger shift for the same velocity. Higher-frequency sound is also attenuated more in tissue, reducing penetration.

  171. Question 171Instrumentation, Quality Assurance and Specialized Vascular Testing

    Excessive color gain on a duplex image most commonly produces:

    1. increased color frame rate
    2. loss of color signal within the vessel lumen
    3. reversal of color flow direction
    4. color blooming into adjacent tissue Correct answer.

    Explanation

    When color gain is set too high, low-level noise is assigned color and spills outside the vessel, producing color blooming. Reducing gain until color just fills the lumen without extending into tissue is the correct adjustment.

  172. Question 172Instrumentation, Quality Assurance and Specialized Vascular Testing

    On a linear array transducer, steering the color box is used to:

    1. optimize the Doppler angle Correct answer.
    2. increase the frame rate
    3. eliminate wall motion artifact
    4. reduce the risk of aliasing

    Explanation

    Steering the color beam changes the angle between the ultrasound beam and the vessel, allowing a more favorable Doppler angle when the vessel runs parallel to the skin surface. This improves the detected Doppler shift.

  173. Question 173Instrumentation, Quality Assurance and Specialized Vascular Testing

    Increasing the dynamic range of a grayscale ultrasound display will:

    1. display fewer shades of gray and increase contrast
    2. reduce the axial resolution
    3. show more gray shades with lower contrast Correct answer.
    4. increase the color frame rate

    Explanation

    Dynamic range determines the range of echo amplitudes that are displayed as shades of gray. A higher dynamic range displays more gray levels, producing a softer image with lower contrast.

  174. Question 174Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which change will increase the color Doppler frame rate?

    1. Increasing the packet size
    2. Increasing the depth of the color box
    3. Reducing the width of the color box Correct answer.
    4. Increasing the color gain

    Explanation

    Frame rate depends on how many color lines and pulses must be processed per frame. Narrowing the color box reduces the number of lines, allowing more frames per second.

  175. Question 175Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which statement describes power Doppler imaging?

    1. It displays flow direction with a color map
    2. It is angle-independent and detects slow flow Correct answer.
    3. It measures peak systolic velocity accurately
    4. It requires a Doppler angle of exactly 60 degrees

    Explanation

    Power Doppler displays the amplitude of the Doppler signal rather than its frequency shift, making it relatively independent of the Doppler angle and sensitive to slow flow. Because direction is not encoded, it cannot show flow direction.

  176. Question 176Instrumentation, Quality Assurance and Specialized Vascular Testing

    A larger sample volume in pulsed wave Doppler will typically produce which spectral change?

    1. Greater spectral broadening Correct answer.
    2. A narrower velocity spectrum
    3. A lower Nyquist limit
    4. A higher peak systolic velocity

    Explanation

    A longer sample volume encompasses a wider range of flow velocities across the vessel profile. The resulting spectral display shows a broader range of frequencies, known as spectral broadening.

  177. Question 177Instrumentation, Quality Assurance and Specialized Vascular Testing

    When measuring the ankle-brachial index, the ankle cuff should be inflated to:

    1. exactly 200 mmHg in every patient regardless
    2. 20 mmHg above estimated systolic pressure Correct answer.
    3. the patient's diastolic pressure measured at rest
    4. 100 mmHg regardless of the patient's pressure

    Explanation

    The cuff is inflated until the Doppler signal disappears and then approximately 20 mmHg higher to ensure arterial occlusion before deflation. This avoids falsely low pressure readings from incomplete occlusion.

  178. Question 178Instrumentation, Quality Assurance and Specialized Vascular Testing

    During segmental pressure testing, a pressure drop of more than 20 mmHg between adjacent cuffs indicates:

    1. a significant stenosis between cuffs Correct answer.
    2. an improperly placed cuff on the limb
    3. normal arterial flow with no stenosis
    4. venous insufficiency with reflux

    Explanation

    A pressure gradient between adjacent segments reflects resistance to flow, and a drop greater than 20 mmHg is generally considered hemodynamically significant. It localizes the level of arterial stenosis or occlusion.

  179. Question 179Instrumentation, Quality Assurance and Specialized Vascular Testing

    For duplex evaluation of lower extremity venous insufficiency, significant reflux in a deep vein is defined as reflux lasting longer than:

    1. 1.0 second Correct answer.
    2. 0.3 second
    3. 3.0 seconds
    4. 0.1 second

    Explanation

    Reflux duration is measured after distal augmentation or Valsalva release. In deep veins, reflux lasting more than 1.0 second is considered significant, whereas superficial veins use a shorter threshold.

  180. Question 180Instrumentation, Quality Assurance and Specialized Vascular Testing

    In duplex surveillance of a brachial artery-to-axillary vein hemodialysis access, a flow volume below which value suggests a failing access?

    1. 1500 mL/min
    2. 1200 mL/min
    3. 800 mL/min
    4. 500 mL/min Correct answer.

    Explanation

    Access flow volume is measured with duplex and a value below approximately 500 mL/min is associated with increased risk of thrombosis and access failure. Serial trends are also important, as a declining flow rate may indicate stenosis.

  181. Question 181Instrumentation, Quality Assurance and Specialized Vascular Testing

    Transcranial Doppler insonation through the temporal window most commonly evaluates which vessel?

    1. Middle cerebral artery Correct answer.
    2. Internal carotid artery at the bifurcation
    3. Vertebral artery at V1
    4. Basilar artery

    Explanation

    The temporal acoustic window lies above the zygomatic arch and provides access to the terminal internal carotid artery and the middle, anterior, and posterior cerebral arteries. The middle cerebral artery is the vessel most commonly interrogated through this window.

  182. Question 182Instrumentation, Quality Assurance and Specialized Vascular Testing

    Photoplethysmography (PPG) of the digits is most useful for:

    1. measuring intracranial blood velocity
    2. assessing digital arterial perfusion Correct answer.
    3. measuring venous reflux time in the saphenous vein
    4. calculating the ankle-brachial index

    Explanation

    PPG detects pulsatile changes in skin blood volume, so a digit waveform reflects arterial perfusion to the finger or toe. It is often used with cold challenge to evaluate vasospastic disorders.

  183. Question 183Instrumentation, Quality Assurance and Specialized Vascular Testing

    A spectral Doppler system operates at a pulse repetition frequency of 6 kHz. What is the Nyquist limit?

    1. 12 kHz
    2. 1.5 kHz
    3. 3 kHz Correct answer.
    4. 6 kHz

    Explanation

    The Nyquist limit is half the pulse repetition frequency. For a PRF of 6 kHz, the limit is 3 kHz, and Doppler shifts above this value will alias.

  184. Question 184Instrumentation, Quality Assurance and Specialized Vascular Testing

    A 5 MHz transducer insonates a vessel at a 60-degree Doppler angle. The detected Doppler shift is 2.0 kHz. Using c = 1540 m/s, what is the blood velocity?

    1. 61.6 cm/s Correct answer.
    2. 46.2 cm/s
    3. 30.8 cm/s
    4. 123 cm/s

    Explanation

    The Doppler equation gives velocity as (shift × c) divided by (2 × transmitted frequency × cos θ). Substituting 2000 Hz, 1540 m/s, 5 × 10^6 Hz, and cos 60 degrees (0.5) yields approximately 0.616 m/s, or 61.6 cm/s.

  185. Question 185Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which quality assurance test uses a string phantom?

    1. testing of electrical safety grounding
    2. assessment of monitor luminance
    3. Doppler velocity accuracy Correct answer.
    4. measurement of acoustic power output

    Explanation

    A string phantom moves a target at a known speed, producing a predictable Doppler shift that can be compared with the displayed velocity. This verifies the accuracy of the system's velocity measurement.

  186. Question 186Instrumentation, Quality Assurance and Specialized Vascular Testing

    What is the primary function of the wall filter in spectral Doppler?

    1. Rejects high-amplitude, low-frequency wall motion signals Correct answer.
    2. Averages consecutive spectral lines to reduce random noise in the display
    3. Eliminates echoes that arise from tissue deep to the sample gate
    4. Automatically reverses the spectrum when flow direction changes

    Explanation

    The wall filter is a high-pass filter that removes the strong, low-frequency Doppler shifts produced by moving vessel walls and valve leaflets, which would otherwise obscure low-velocity flow near the baseline. Setting it too high can erase genuine slow flow such as venous or diastolic flow.

  187. Question 187Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which statement best describes continuous wave Doppler?

    1. It transmits and receives continuously, so signals from the whole beam path are superimposed. Correct answer.
    2. It is limited to velocities below the Nyquist limit of the transmitted frequency.
    3. It provides range resolution along the entire beam path.
    4. It requires a sample gate to be positioned over the vessel of interest.

    Explanation

    Continuous wave Doppler uses separate transmitting and receiving elements operating simultaneously, so there is no depth discrimination. Because pulses are not used, there is effectively no pulse repetition frequency limit and therefore no aliasing, which makes it useful for very high velocities.

  188. Question 188Instrumentation, Quality Assurance and Specialized Vascular Testing

    During a duplex examination, the Doppler beam is oriented at 90 degrees to the axis of the vessel. What is expected?

    1. No detectable Doppler shift Correct answer.
    2. Aliasing of the spectral waveform
    3. A maximal Doppler shift with spectral broadening
    4. A measured velocity approximately double the true value

    Explanation

    The Doppler shift is proportional to the cosine of the angle between the beam and the direction of flow. At 90 degrees the cosine is zero, so no Doppler shift is generated and no velocity can be measured. Angles of 60 degrees or less are preferred to keep the cosine correction accurate.

  189. Question 189Instrumentation, Quality Assurance and Specialized Vascular Testing

    What is the purpose of time gain compensation?

    1. Increase the acoustic output power progressively with increasing depth of penetration
    2. Compensate for attenuation so similar reflectors appear equally bright at all depths Correct answer.
    3. Allow the Doppler gain to be adjusted independently of the two-dimensional image gain
    4. Reduce the width of the color box so that the color frame rate increases

    Explanation

    Time gain compensation applies increasing amplification to echoes that arrive later, offsetting the greater attenuation suffered by sound traveling to and from deeper tissues. Correctly set, identical reflectors at different depths are displayed with similar brightness.

  190. Question 190Instrumentation, Quality Assurance and Specialized Vascular Testing

    A tissue-equivalent phantom contains pairs of thin filaments spaced at progressively smaller intervals measured along the beam axis. Determining the smallest separation at which the two filaments remain distinct tests which parameter?

    1. Elevational resolution
    2. Axial resolution Correct answer.
    3. Color Doppler sensitivity
    4. Lateral resolution

    Explanation

    Axial resolution is the ability to separate two reflectors lying along the path of the beam and depends on spatial pulse length, which is shortened by higher frequency and fewer cycles per pulse. Measurements made across the beam width or in the elevational plane test other resolution dimensions.

  191. Question 191Instrumentation, Quality Assurance and Specialized Vascular Testing

    The mechanical index displayed on an ultrasound system is most closely associated with which potential bioeffect?

    1. Radiation-force displacement of erythrocytes
    2. Heating of the transducer crystal during prolonged scanning
    3. Cavitation related to peak negative pressure Correct answer.
    4. Tissue heating from absorbed acoustic energy

    Explanation

    Mechanical index estimates the likelihood of cavitation and is calculated from the peak rarefactional pressure divided by the square root of the transmitted frequency. Non-mechanical effects such as tissue heating are instead reflected by the thermal index.

  192. Question 192Instrumentation, Quality Assurance and Specialized Vascular Testing

    Tissue harmonic imaging improves image quality primarily because the received harmonic signal:

    1. Contains less side-lobe and clutter artifact than the fundamental signal Correct answer.
    2. Removes the need for time gain compensation
    3. Is generated only by moving blood within the vessel lumen
    4. Has a higher frequency and therefore penetrates more deeply

    Explanation

    Harmonic energy builds up preferentially along the main beam, so weak harmonic signals returning from side lobes and body-wall clutter are suppressed relative to the fundamental. The result is improved contrast resolution and reduced artifact, at the cost of somewhat reduced penetration.

  193. Question 193Instrumentation, Quality Assurance and Specialized Vascular Testing

    In pulsed wave Doppler, what determines the highest pulse repetition frequency that can be used without range ambiguity?

    1. The wall filter setting
    2. The transmitted Doppler frequency
    3. The depth of the sample volume Correct answer.
    4. The size of the sample volume

    Explanation

    Each pulse must complete its round trip to the sample depth and back before the next pulse is emitted, so the usable PRF is limited by the speed of sound divided by twice the depth. Selecting a deeper gate therefore lowers the maximum PRF and predisposes the tracing to aliasing.

  194. Question 194Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which transducer is most appropriate for imaging the superficial femoral artery in a slender adult?

    1. A 2 to 5 MHz curved array
    2. A 5 to 10 MHz linear array Correct answer.
    3. A 1 to 3 MHz phased array
    4. A 3.5 MHz continuous wave pencil probe

    Explanation

    A high-frequency linear array provides the spatial resolution needed for superficial vessels and a rectangular field of view that matches the straight course of the superficial femoral artery. Lower-frequency curved and phased arrays trade resolution for penetration and are reserved for deeper structures.

  195. Question 195Instrumentation, Quality Assurance and Specialized Vascular Testing

    A patient has a normal resting ankle-brachial index. After treadmill exercise the ankle pressure falls by 25 mmHg and the patient reports calf pain. This response indicates:

    1. A hemodynamically significant arterial stenosis in the symptomatic limb Correct answer.
    2. Inadequate cuff placement during the post-exercise measurement
    3. Venous claudication from outflow obstruction
    4. Normal exercise physiology with no arterial disease

    Explanation

    Normally ankle pressure rises slightly or remains unchanged with exercise because peripheral resistance falls. A post-exercise fall of about 15 to 20 mmHg or more, especially with reproducible symptoms, indicates a flow-limiting arterial lesion and helps localize the symptomatic level.

  196. Question 196Instrumentation, Quality Assurance and Specialized Vascular Testing

    Which set of findings describes a well-functioning native arteriovenous fistula according to the commonly cited 'rule of six'?

    1. Flow volume above 1000 mL/min, diameter above 10 mm, depth less than 4 mm
    2. Flow volume above 300 mL/min, diameter above 4 mm, depth below 8 mm
    3. Flow volume above 100 mL/min, diameter above 3 mm, depth below 10 mm
    4. Flow volume above 600 mL/min, diameter above 6 mm, depth less than 6 mm Correct answer.

    Explanation

    The rule of sixes describes a mature, functional fistula as having a flow volume greater than 600 mL/min, a diameter of at least 6 mm, and a depth of 6 mm or less from the skin. Values well outside these ranges prompt closer evaluation for stenosis, immaturity, or excessive depth.

  197. Question 197Instrumentation, Quality Assurance and Specialized Vascular Testing

    The color velocity scale of a duplex system is set far higher than the velocities present in a scanned vein. What is the most likely result?

    1. Color appears outside the vessel wall
    2. The vein fills with aliased color
    3. Color write priority increases
    4. The vein shows little or no color filling Correct answer.

    Explanation

    When the color scale greatly exceeds the actual Doppler shifts, the autocorrelator detects little or no mean shift and the vessel is not colorized. The scale should be lowered to match the low velocities of venous flow.

  198. Question 198Instrumentation, Quality Assurance and Specialized Vascular Testing

    During penile duplex after intracavernosal vasodilator injection, the peak systolic velocity in a cavernosal artery measures 20 cm/s. This finding is most consistent with:

    1. Venous leak from veno-occlusive dysfunction
    2. Arterial insufficiency Correct answer.
    3. Normal cavernosal arterial inflow
    4. High-flow priapism

    Explanation

    A normal cavernosal artery peak systolic velocity after vasodilator stimulation is generally above 30 to 35 cm/s, and values below roughly 25 to 30 cm/s suggest arterial inflow insufficiency. Venous occlusive dysfunction is instead assessed from the end-diastolic velocity, which remains elevated.

  199. Question 199Instrumentation, Quality Assurance and Specialized Vascular Testing

    Transcranial Doppler of the middle cerebral artery shows an elevated pulsatility index with a normal mean velocity. This pattern most suggests:

    1. A proximal high-grade stenosis of the insonated segment
    2. Increased distal cerebrovascular resistance Correct answer.
    3. Vasospasm of the middle cerebral artery
    4. An inadequate temporal acoustic window

    Explanation

    The pulsatility index reflects the resistance of the vascular bed downstream from the insonated segment, so an elevated index with preserved mean velocity points to raised distal resistance. Vasospasm, by contrast, produces a focal rise in velocity rather than an isolated rise in pulsatility.

  200. Question 200Instrumentation, Quality Assurance and Specialized Vascular Testing

    A spectral Doppler tracing shows a second, fainter waveform mirroring the true spectrum on the opposite side of the baseline. This is most likely caused by:

    1. Aliasing caused by a pulse repetition frequency that is set too low
    2. Electrical interference from nearby equipment in the room
    3. A wall filter set so high that low-velocity signals disappear
    4. A strong reflector adjacent to the vessel reflecting the beam Correct answer.

    Explanation

    A mirror image occurs when a strong specular reflector lying close to the vessel at nearly 90 degrees to the beam redirects the Doppler signal, displaying a duplicate spectrum on the opposite side of the baseline. Adjusting the Doppler angle or repositioning the transducer usually removes it.

RVT flashcards

32 cards. Read the term and the definition here. The deck at the top of this page is optional.

ICA/CCA velocity ratio
Peak systolic velocity in the internal carotid artery divided by peak systolic velocity in the distal common carotid artery; a value of 2.0 or greater supports a 50% or greater ICA stenosis.
NASCET stenosis formula
Percent stenosis equals one minus the residual lumen diameter divided by the normal distal internal carotid diameter, multiplied by 100; it is the angiographic standard used to validate duplex criteria.
Ophthalmic artery flow reversal
Reversed ophthalmic artery flow on orbital Doppler is an indirect sign of hemodynamically significant ipsilateral internal carotid stenosis or occlusion, with collateral supply reaching the orbit from the contralateral side.
Subclavian steal flow reversal
Retrograde vertebral artery flow, often provoked by ipsilateral arm exercise, occurs when a proximal subclavian stenosis or occlusion reverses the normal vertebrobasilar pressure gradient.
Normal internal carotid waveform
A low-resistance spectral pattern with a broad systolic peak and continuous antegrade diastolic flow, reflecting the low-resistance cerebral vascular bed served by the internal carotid artery.
String sign near occlusion
A thin, thread-like residual color and spectral flow channel in a nearly occluded internal carotid artery; distal velocities may fall rather than rise, so stenosis grading by velocity alone fails.
Ankle-brachial index formula
Highest ankle systolic pressure from the dorsalis pedis or posterior tibial artery divided by the higher of the two brachial systolic pressures, using the higher-pressure lower extremity value.
ABI interpretation thresholds
Values above 1.30 indicate non-compressible calcified arteries, 1.00 to 1.30 is normal, 0.91 to 0.99 is borderline, and 0.90 or less defines peripheral arterial disease.
Triphasic waveform significance
Three-phase flow, with sharp systolic forward flow, early diastolic reversal, and late diastolic forward flow, indicates a normal compliant conduit without an upstream hemodynamically significant stenosis.
Segmental pressure gradient
A pressure drop greater than 20 to 30 mmHg between adjacent thigh, calf, and ankle cuffs localizes a hemodynamically significant arterial stenosis to the segment between those cuffs.
Toe-brachial index
Toe systolic pressure divided by brachial systolic pressure; normal exceeds 0.70, and the index is preferred over ankle pressure when medial calcinosis makes ankle vessels non-compressible.
Post-exercise ABI
Ankle-brachial index measured after treadmill walking; a fall of more than 20 percent from baseline, or a delayed return to baseline, indicates hemodynamically significant occult disease.
Venous spontaneity and phasicity
Normal limb vein Doppler shows spontaneous flow that varies with respiration; continuous non-phasic flow suggests obstruction proximal to the sample site, while absent spontaneous flow suggests occlusion at that site.
Venous reflux thresholds
Reflux lasting more than 0.5 second defines incompetence in superficial and perforating veins, while more than 1.0 second is required in the common femoral and popliteal deep veins.
Acute versus chronic thrombus
Acute thrombus is hypoechoic with a distended, non-compressible vein; chronic thrombus becomes echogenic, the vein contracts, and collaterals develop, often leaving residual valvular reflux.
Valsalva response in femoral vein
A normal Valsalva maneuver abolishes or briefly reverses femoral venous flow; failure to do so, or continuous flow, implies a patent pathway around a proximal obstruction.
Venous augmentation test
Manual distal compression should sharply increase venous velocity at the sampled site; blunted or absent augmentation indicates obstruction between the compression point and the Doppler sample volume.
Aortic aneurysm size criteria
An infrarenal aortic diameter of 3.0 cm or more defines an aneurysm; elective repair is generally considered at 5.5 cm in men and 5.0 cm in women, or with rapid growth.
Renal-aortic ratio
Peak systolic velocity in the renal artery divided by peak systolic velocity in the abdominal aorta; a ratio of 3.5 or greater supports a 60 percent or greater renal artery stenosis.
Tardus-parvus waveform
Dampened downstream flow with a slow systolic upstroke and low peak velocity, produced by a hemodynamically significant stenosis or occlusion proximal to the sampling location.
Hepatic artery resistive index
Normal hepatic arterial resistive index is about 0.55 to 0.70; a value below 0.5 after transplantation raises concern for hepatic artery stenosis, thrombosis, or outflow abnormality.
Portal hypertension duplex signs
Main portal vein diameter above 13 mm with mean velocity below roughly 15 cm per second, plus loss or reversal of hepatopetal flow, suggests portal hypertension.
Doppler equation
Doppler shift equals two times the transmitted frequency times blood velocity times the cosine of the Doppler angle, divided by the speed of sound in tissue of about 1540 m/s.
Nyquist limit
The highest frequency shift a pulsed Doppler system can detect without ambiguity, equal to half the pulse repetition frequency; velocities exceeding it produce aliasing.
Aliasing correction options
Increase the pulse repetition frequency or velocity scale, lower the transducer frequency, adjust the baseline, increase the Doppler angle toward 60 degrees, or switch to continuous wave Doppler.
Wall filter function
A high-pass filter that removes low-frequency, high-amplitude echoes from vessel wall and valve motion; set too high it also removes slow venous or post-stenotic flow.
Power Doppler characteristics
Uses Doppler amplitude rather than frequency shift, so it is angle-independent, does not alias, and detects slow flow, but gives no direction or velocity information.
Lindegaard ratio
Mean flow velocity in the middle cerebral artery divided by velocity in the ipsilateral extracranial internal carotid artery; a ratio above 3.0 supports vasospasm rather than hyperemia.
Transcranial Doppler vasospasm threshold
Middle cerebral artery mean velocity above 120 cm per second suggests vasospasm, and is most meaningful when compared with the Lindegaard ratio and serial baseline values.
Treadmill exercise protocol
Walk at 2 miles per hour on a 12 percent grade for up to 5 minutes, with ankle-brachial indices measured before and immediately after exercise to provoke occult disease.
Flow phantom velocity accuracy
Velocity measured in a calibrated Doppler flow phantom should agree with the true value within about 10 percent; larger deviations indicate angle, scale, or transducer calibration error.
Dialysis access flow threshold
A native fistula or graft with access flow below 500 to 600 mL per minute, or a stenosis with a peak systolic velocity ratio above 2, suggests significant stenosis.

RVT study guide

Cerebrovascular

Two questions drive most items

Nearly every cerebrovascular item asks one of two things: which vessel am I looking at, and how tight is the narrowing? Answer the first before you grade the second.

Telling ICA from ECA

The ICA is usually larger, sits posterolateral, has no cervical branches, and shows a low-resistance waveform with continuous forward diastolic flow because it supplies the brain. The ECA is smaller, anteromedial, gives off branches (superior thyroid first), and shows a high-resistance, triphasic pattern with early diastolic reversal. A temporal tap or jaw movement confirms the ECA; the ICA does not respond to either. Trap: a severely stenotic ICA can lose diastolic flow and mimic the ECA, so find the bifurcation and confirm anatomy before you trust the waveform.

Grading ICA stenosis

Widely used consensus criteria combine four parameters: peak systolic velocity (PSV), end-diastolic velocity (EDV), the ICA/CCA PSV ratio, and plaque plus spectral broadening. Approximate thresholds: normal, PSV under 125 cm/s and ratio under 2.0; 50 to 69 percent, PSV 125 to 230, EDV 40 to 100, ratio 2.0 to 4.0; 70 percent or greater, PSV above 230, EDV above 100, ratio above 4.0. Sample the CCA from the mid-to-distal segment, well below the bulb, or the ratio becomes meaningless. Near-occlusion may show a string sign or paradoxically low velocities; do not call it occluded on velocity alone.

Worked example

ICA PSV 260 cm/s, EDV 110 cm/s, distal CCA PSV 60 cm/s. Ratio = 260 divided by 60 = 4.3. With PSV above 230, EDV above 100 and ratio above 4.0, this meets criteria for a 70 percent or greater diameter-reducing stenosis.

Peripheral Arterial

Physiologic, not anatomic

Segmental pressures, plethysmography and Doppler waveform analysis are physiologic tests. They prove a lesion exists and roughly localize it; they do not depict the plaque.

ABI technique and pitfalls

Use a cuff about 20 percent wider than the limb; a cuff that is too narrow falsely elevates pressure. Rest the patient supine for about 10 minutes, measure both brachial pressures and use the higher, then measure dorsalis pedis and posterior tibial at each ankle and use the higher pressure per ankle. ABI equals highest ankle pressure divided by highest brachial pressure. Interpretation: above 1.4 is non-compressible (medial calcinosis in diabetes or end-stage renal disease); 1.0 to 1.4 normal; 0.9 to 1.0 borderline; 0.5 to 0.9 typical of claudication; below 0.4 suggests rest pain or tissue loss. Trap: a normal-looking ABI in a diabetic patient with rest pain. Calcified vessels invalidate the ratio, so go to toe pressures or waveforms.

Waveforms and velocity

Normal lower-extremity Doppler is triphasic: sharp upstroke, brief early reversal, small forward diastolic component. Proximal to a stenosis the waveform dampens and becomes monophasic. At the stenosis the PSV roughly doubles with spectral broadening; downstream flow is turbulent. A systolic pressure drop of 20 to 30 mmHg between adjacent segments is significant.

Worked example

ABI of 1.5 in a patient with rest pain is not reassurance. Obtain toe pressures (normal above 60 mmHg, toe-brachial index above 0.7) or plethysmographic waveforms to evaluate the truly compressible distal bed.

Peripheral Venous

The normal venous signature

A normal lower-extremity vein is spontaneous, phasic with respiration, augments with distal compression and release, non-pulsatile, and fully compressible under probe pressure. Loss of phasicity points to proximal obstruction; continuous non-phasic flow may indicate a proximal stenosis or collateral pathway.

Acute versus chronic thrombus

Acute deep vein thrombosis: the vein is distended, walls are thin, the lumen is hypoechoic or anechoic, it will not compress, and collaterals are absent. Chronic changes: small echogenic vein, partial compressibility, recanalization, prominent collaterals, damaged valves. Trap: partially recanalized chronic thrombus is often misread as acute. Vein caliber and echogenicity, not compression alone, are the discriminators.

Reflux testing

Test with the patient upright or in reverse Trendelenburg. Use Valsalva for the common femoral segment and manual or cuff compression more distally. Reflux is defined by duration of reversed flow: roughly above 0.5 s in deep veins and in the great and small saphenous veins, and above 0.35 s in perforating veins, though thresholds vary by laboratory protocol. Superficial reflux at the saphenofemoral junction above 0.5 s with a dilated great saphenous vein supports incompetence.

Worked example

Calf compression with release produces 1.2 s of retrograde popliteal flow. That exceeds the 0.5 s threshold, so the popliteal valve is incompetent. Note that the maneuver matters: reflux measured with the patient supine and no provocation is frequently underestimated.

Abdominal and Visceral Vascular

Aortic measurement

Measure an aneurysm outer wall to outer wall, anteroposterior, in a plane perpendicular to the vessel axis. A tortuous aorta measured obliquely is falsely enlarged, a classic distractor. Infrarenal aorta above 3 cm is aneurysmal. Report maximal diameter, extent, renal and iliac involvement, and mural thrombus; measure the outer wall, never the flow lumen. Shorten surveillance intervals as size increases; rapid growth, roughly 0.5 cm in six months or 1 cm per year, prompts referral. After endovascular repair, look for endoleak, sac enlargement and graft limb flow.

Renal and mesenteric arteries

Normal renal artery PSV is below roughly 180 to 200 cm/s; a renal-aortic ratio above 3.5 supports significant stenosis. Kidney length is normally 9 to 12 cm; a small kidney with resistive index above 0.7 suggests parenchymal disease rather than correctable stenosis. For mesenteric ischemia, evaluate the celiac axis and superior mesenteric artery in sagittal and axial views. The SMA is high resistance when fasting and becomes low resistance after eating, so a postprandial study can mask or mimic disease.

Worked example

A 4.2 cm infrarenal aorta measured at a steep angle through a tortuous segment may truly be 3.5 cm. Re-measure in two planes and perpendicular to the axis before you commit to a diameter, and always state the measurement method in the report.

Instrumentation, Quality Assurance and Specialized Vascular Testing

Instrumentation

The Doppler equation

Shift equals (2 times transmitted frequency times velocity times cosine of the Doppler angle) divided by the speed of sound in tissue. The shift rises with transmitted frequency and blood velocity, and falls as the angle approaches 90 degrees. At 60 degrees you record half the maximum shift with the least error from small angle mistakes; at 90 degrees you record nothing. Rule: keep the angle at or below 60 degrees and angle-correct only in a steerable, sagittal-type plane. In the transverse plane no meaningful angle exists and velocity should not be reported.

Frequency trade-offs

Higher frequency gives better axial resolution but poorer penetration; lower frequency penetrates at the cost of resolution. Choose the highest frequency that reaches the depth of interest. Raising Doppler frequency also raises the shift and can push you into aliasing.

Aliasing and system controls

Nyquist limit equals PRF divided by 2. A shift beyond that limit wraps to the opposite side of the baseline. Fix it by increasing PRF or scale, lowering transmitted frequency, increasing the Doppler angle, or shifting the baseline. The wall filter removes high-amplitude, low-velocity clutter from wall motion, but too much filter erases slow flow. Excess color gain produces bleed outside the vessel; insufficient gain produces dropout.

Worked example

Color fills the vessel but the spectral trace wraps around the scale. Raise the velocity scale (PRF) first. If the vessel is deep and the signal is weak, drop to a lower-frequency transducer rather than increasing PRF further.

Quality Assurance and Specialized Vascular Testing

Why QA appears on the exam

Quality items test whether you accept that a velocity is only as trustworthy as the machine calibration and your technique. A Doppler flow phantom verifies velocity accuracy, axial and spatial resolution, penetration and sensitivity. Tissue-equivalent phantoms and gray-scale targets assess image performance, and flow phantoms assess color sensitivity. Keep logs, follow a written schedule, and document corrective action. Common traps: confusing sensitivity, the ability to detect low flow, with resolution, the ability to separate two structures; and assuming a new transducer needs no testing.

Specialized tests worth knowing

Transcutaneous oximetry: values below about 30 mmHg predict poor healing, values above about 40 mmHg are generally adequate for amputation healing. Temperature, edema and probe placement all affect the reading, so include a reference site. Digital and segmental plethysmography with toe pressures is used when arteries are non-compressible; toe pressure above 60 mmHg and toe-brachial index above 0.7 are normal. Cold challenge compares baseline digital pressures or plethysmography with values after cold immersion, used in Raynaud evaluation. Dialysis access surveillance uses the rule of sixes: flow above 600 mL/min, diameter above 6 mm, depth under 6 mm, plus a focal velocity step-up with a PSV ratio near 2 to 3 to 1 suggesting stenosis.

Worked example

A forearm arteriovenous fistula with brachial artery flow of 350 mL/min and a 3 to 1 PSV step-up at the venous anastomosis indicates a flow-limiting stenosis. That combination warrants referral rather than routine follow-up.

RVT cheat sheet

Quick reference

Cerebrovascular

  • An ICA peak systolic velocity of 125 cm/s or greater together with an ICA/CCA velocity ratio of 2.0 or greater identifies a 50% or greater diameter stenosis under the SRU consensus criteria.

  • NASCET stenosis (%) = (1 - residual lumen diameter / normal distal ICA diameter) x 100.

  • ICA/CCA ratio = peak systolic velocity in the ICA divided by peak systolic velocity in the distal (mid) common carotid artery.

  • A normal ICA spectral waveform is low resistance, with a broad systolic peak and continuous antegrade diastolic flow.

  • PSV of 230 cm/s or greater, or end diastolic velocity of 100 cm/s or greater, in the ICA suggests 80% or greater stenosis; a thread-like residual channel indicates near occlusion.

  • Reversed ophthalmic artery flow is an indirect sign of hemodynamically significant ipsilateral ICA stenosis or occlusion.

  • Retrograde vertebral artery flow, especially after ipsilateral arm exercise, supports subclavian steal.

  • Echolucent plaque carries a higher embolic risk than densely calcified plaque; plaque surface ulceration is likewise associated with increased stroke risk.

Quick reference

Peripheral Arterial

  • ABI = highest ankle systolic pressure (dorsalis pedis or posterior tibial) divided by the higher of the two brachial systolic pressures.

  • ABI interpretation: above 1.30 non-compressible; 1.00-1.30 normal; 0.91-0.99 borderline; 0.90 or less abnormal.

  • A triphasic waveform is normal in limb arteries; monophasic, dampened flow indicates a proximal stenosis or collateral-dependent segment.

  • A segmental pressure gradient greater than 20-30 mmHg between adjacent cuffs localizes a hemodynamically significant lesion to that segment.

  • A peak systolic velocity ratio of 2.0 or greater across a stenosis corresponds to roughly 50% or greater diameter reduction, with higher ratios indicating tighter lesions.

  • Toe-brachial index above 0.70 is normal and is preferred when medial calcinosis makes ankle pressures falsely elevated.

  • Poiseuille relationship: volume flow varies with the fourth power of radius, so a 50% diameter reduction lowers flow roughly sixteen-fold at constant pressure.

  • A fall in ABI greater than 20% from baseline after treadmill exercise indicates hemodynamically significant occult disease.

Quick reference

Peripheral Venous

  • Normal venous Doppler is spontaneous, phasic with respiration, augments with distal compression, and ceases or briefly reverses with Valsalva.

  • Deep vein thrombosis criteria: non-compressibility, intraluminal echogenicity, absent spontaneous flow, and loss of respiratory phasicity.

  • Reflux thresholds: greater than 0.5 second in superficial and perforating veins; greater than 1.0 second in common femoral and popliteal veins.

  • Acute thrombus is hypoechoic with a distended vein; chronic thrombus is echogenic, the vein is small, and collaterals have formed.

  • Continuous, non-phasic femoral venous flow suggests obstruction proximal to the sampling site.

  • Blunted or absent augmentation after distal compression indicates obstruction between the compression point and the sample volume.

  • Reflux in the great saphenous vein is assessed with the patient standing and the limb dependent, sampling at the saphenofemoral junction and along the thigh.

  • Calf veins are paired (posterior tibial and peroneal); soleal and gastrocnemius muscular sinuses are not paired and require careful compressibility assessment.

Quick reference

Abdominal and Visceral Vascular

  • Infrarenal aortic diameter of 3.0 cm or more defines an aneurysm; elective repair is generally considered at 5.5 cm in men and 5.0 cm in women, or with growth over about 1 cm per year.

  • Renal artery PSV below 180 cm/s is normal; 180-200 cm/s or greater with a renal-aortic ratio of 3.5 or more suggests 60% or greater stenosis.

  • Renal-aortic ratio = renal artery peak systolic velocity divided by abdominal aortic peak systolic velocity.

  • A tardus-parvus waveform (slow upstroke, low peak velocity) distal to a stenosis indicates hemodynamically significant obstruction.

  • Normal hepatic artery resistive index is approximately 0.55-0.70; an index below 0.5 after transplantation raises concern for stenosis or thrombosis.

  • Portal hypertension is suggested by main portal vein diameter above 13 mm with mean velocity below about 15 cm/s (normal 15-30 cm/s).

  • Superior mesenteric artery PSV above 275 cm/s suggests 70% or greater stenosis; the normal fasting SMA waveform is high resistance with low diastolic flow.

  • Endograft surveillance evaluates sac diameter, endoleaks, and graft limb stenosis; any perfused endoleak with sac enlargement requires intervention planning.

Quick reference

Instrumentation

  • Doppler equation: frequency shift = (2 x transmitted frequency x velocity x cosine of Doppler angle) / speed of sound in tissue (about 1540 m/s).

  • Maintain a Doppler angle of 60 degrees or less, ideally 45-60 degrees, because cosine error grows rapidly as the angle approaches 90 degrees.

  • Nyquist limit = pulse repetition frequency / 2; velocities above this value alias.

  • Aliasing corrections: increase pulse repetition frequency or scale, lower transducer frequency, adjust baseline, increase Doppler angle, or switch to continuous wave Doppler.

  • Continuous wave Doppler has no depth resolution; pulsed Doppler with a sample volume provides range gating.

  • A higher transmitted frequency improves spatial resolution but increases attenuation and reduces penetration depth.

  • The wall filter is a high-pass filter; set too high it removes low-velocity flow such as venous or post-stenotic flow.

  • Power Doppler is angle-independent, does not alias, and detects slow flow, but provides no direction or velocity information.

Quick reference

Quality Assurance and Specialized Vascular Testing

  • Routine QA includes daily image uniformity checks, vertical and horizontal distance accuracy within about 2% or 1 mm, and an annual comprehensive system evaluation.

  • Velocity measured in a calibrated flow phantom should agree with the true value within approximately 10%.

  • Tissue-mimicking phantoms assess dead zone, axial and lateral resolution, penetration depth, and gray-scale uniformity.

  • Lindegaard ratio = middle cerebral artery mean velocity divided by ipsilateral extracranial internal carotid velocity; above 3.0 indicates vasospasm rather than hyperemia.

  • Transcranial Doppler: middle cerebral artery mean velocity above 120 cm/s suggests vasospasm; normal mean velocity is roughly 40-70 cm/s.

  • Treadmill protocol: 2 mph at 12% grade for up to 5 minutes, with ankle-brachial indices obtained before and after walking.

  • Dialysis access surveillance: access flow below 500-600 mL/min or an intra-access peak systolic velocity ratio above 2 suggests significant stenosis.

  • Bioeffects are managed with ALARA, monitoring thermal index and mechanical index, and minimizing dwell time at high output.

RVT key terms

Internal carotid artery (ICA)

A branch of the common carotid bifurcation, usually lying lateral and posterior to the external carotid, that supplies the anterior brain and eye. It has no normal cervical branches and shows a low-resistance waveform.

Related: External carotid artery (ECA), Circle of Willis, Ophthalmic artery flow reversal
External carotid artery (ECA)

The branch of the common carotid that supplies the face, scalp, and neck. It typically gives off cervical branches, lies medial and anterior to the internal carotid, and shows a higher-resistance waveform.

Related: Internal carotid artery (ICA), Ophthalmic artery flow reversal
Circle of Willis

A ring of interconnected arteries at the base of the brain that joins the internal carotid and vertebrobasilar circulations through communicating arteries, allowing collateral flow when one supply route is compromised.

Related: Internal carotid artery (ICA), Vertebral artery, Transcranial Doppler (TCD)
Vertebral artery

A cervical artery, usually arising from the subclavian, that ascends through the transverse foramina of the spine and joins its partner to form the basilar artery, supplying the posterior circulation.

Related: Subclavian steal syndrome, Circle of Willis
Transcranial Doppler (TCD)

A low-frequency Doppler technique, usually about 2 MHz, that insonates intracranial arteries through acoustic windows in the skull to measure velocities in the circle of Willis and its branches.

Related: Circle of Willis, Subclavian steal syndrome
Subclavian steal syndrome

Retrograde vertebral artery flow caused by a significant stenosis or occlusion of the proximal subclavian artery, which reverses the pressure gradient and diverts blood away from the posterior circulation, especially during arm exertion.

Related: Vertebral artery, Transcranial Doppler (TCD)
Ophthalmic artery flow reversal

Reversal of the normal direction of ophthalmic artery flow on orbital Doppler, a finding that suggests severe ipsilateral internal carotid stenosis or occlusion with collateral supply through external carotid branches.

Related: Internal carotid artery (ICA), External carotid artery (ECA)
Spectral broadening

Widening or filling-in of the Doppler spectral waveform caused by a broad range of red cell velocities within the sample volume, typical of post-stenotic turbulence or of an overly large sample gate.

Related: Aliasing, Tardus-parvus waveform
Ankle-brachial index (ABI)

The ratio of the higher ankle systolic pressure, from the dorsalis pedis or posterior tibial artery, to the higher brachial systolic pressure; used to detect and grade lower-extremity arterial obstruction.

Related: Segmental pressures, Toe-brachial index (TBI)
Segmental pressures

Systolic pressures obtained with cuffs at several levels of the limb, such as thigh, calf, and ankle, so that pressure gradients between adjacent segments localize and quantify arterial occlusive disease.

Related: Ankle-brachial index (ABI), Pulse volume recording (PVR)
Pulse volume recording (PVR)

A plethysmographic method that records volume changes in a limb segment during each cardiac cycle using pneumatic cuffs, producing a waveform whose amplitude and contour reflect arterial perfusion.

Related: Segmental pressures, Photoplethysmography (PPG)
Triphasic waveform

A normal peripheral arterial spectral waveform with three phases: forward systolic flow, brief early diastolic reversal, and late diastolic forward flow, reflecting a compliant, high-resistance vascular bed.

Related: Tardus-parvus waveform, Pulse volume recording (PVR)
Tardus-parvus waveform

A damped arterial Doppler waveform with a slow systolic upstroke and a low, rounded peak, seen distal to a severe stenosis or occlusion because of reduced pulsatility and delayed systolic arrival.

Related: Triphasic waveform, Chronic limb-threatening ischemia
Toe-brachial index (TBI)

The ratio of toe systolic pressure to brachial systolic pressure, measured with a small digital cuff and plethysmograph, useful when ankle pressures are falsely elevated by medial calcinosis.

Related: Ankle-brachial index (ABI), Photoplethysmography (PPG)
Exercise treadmill test

A provocative study in which the patient walks on a treadmill and post-exercise ankle pressures are compared with resting values; a fall in pressure indicates hemodynamically significant arterial disease.

Related: Ankle-brachial index (ABI), Segmental pressures
Chronic limb-threatening ischemia

Advanced peripheral arterial disease producing ischemic rest pain, nonhealing ulceration, or gangrene from inadequate limb perfusion, carrying a high risk of limb loss without revascularization.

Related: Ankle-brachial index (ABI), Toe-brachial index (TBI)
Venous reflux

Retrograde venous flow through incompetent valves, provoked by Valsalva or distal compression and timed on duplex; prolonged reflux in deep or superficial veins indicates venous insufficiency.

Related: Valve closure time, Valsalva maneuver
Valsalva maneuver

A forced expiration against a closed glottis that raises intra-abdominal pressure and challenges venous valves; used during duplex to provoke and time reflux in the common femoral and proximal veins.

Related: Venous reflux, Venous augmentation
Venous augmentation

A brief increase in the venous Doppler signal produced by squeezing the limb distal to the sample site; an absent or weak response suggests obstruction between the compression point and the probe.

Related: Deep vein thrombosis (DVT), Venous reflux
Valve closure time

The duration of retrograde flow after a provocative maneuver before the valve closes; commonly used thresholds are greater than about 0.5 seconds in superficial veins and 1 second in deep veins.

Related: Venous reflux, Valsalva maneuver
Deep vein thrombosis (DVT)

Clot formation within the deep veins, most often of the legs, identified on duplex by failure of vein compression, intraluminal echogenic material, and absent or abnormal flow and augmentation.

Related: Venous augmentation, Venous reflux
Photoplethysmography (PPG)

An optical technique that detects changes in skin blood volume using reflected infrared light, applied with calf exercise and tourniquets to measure venous refilling time and assess insufficiency.

Related: Calf muscle pump, Venous reflux
Calf muscle pump

The muscular and valvular mechanism in the calf that propels venous blood toward the heart during walking; its failure contributes to ambulatory venous hypertension and chronic venous disease.

Related: Photoplethysmography (PPG), Venous reflux
Renal-aortic ratio (RAR)

The ratio of peak systolic velocity in the renal artery to that in the adjacent abdominal aorta; a value of about 3.5 or greater supports a hemodynamically significant renal artery stenosis.

Related: Renal artery duplex
Renal artery duplex

Ultrasound evaluation of the renal arteries and kidneys using velocity criteria, waveform analysis, and kidney size to detect stenosis and judge its hemodynamic effect on the kidney.

Related: Renal-aortic ratio (RAR)
Abdominal aortic aneurysm (AAA)

A focal dilation of the abdominal aorta, generally defined as a diameter of 3 cm or more, most often located below the renal arteries, with rupture risk rising as the diameter increases.

Related: Endoleak
Endoleak

Persistent blood flow within an aortic aneurysm sac after endovascular repair, classified by its source such as attachment site, branch vessel, fabric defect, graft porosity, or endotension, and often requiring reintervention.

Related: Abdominal aortic aneurysm (AAA)
Portal hypertension

Abnormally high pressure in the portal venous system, often from cirrhosis, suggested on ultrasound by portal vein dilation, slowed or reversed flow, splenomegaly, ascites, and portosystemic collaterals.

Median arcuate ligament syndrome

Compression of the origin of the celiac artery by the median arcuate ligament of the diaphragm, producing a focal hook-like narrowing with accelerated velocity that typically varies with respiration.

Doppler angle

The angle between the ultrasound beam and the direction of blood flow; velocity calculation requires angle correction, and the cosine of this angle strongly affects the measured frequency shift, ideally keeping it at 60 degrees or less.

Related: Aliasing, Pulse repetition frequency (PRF)
Aliasing

An ambiguous Doppler display that occurs when the frequency shift exceeds the Nyquist limit, causing wrap-around of the spectral waveform or color inversion; corrected by raising PRF, lowering transducer frequency, or adjusting the baseline.

Related: Pulse repetition frequency (PRF), Doppler angle
Pulse repetition frequency (PRF)

The rate at which ultrasound pulses are transmitted, which determines the Nyquist limit and therefore the highest velocity that can be measured without aliasing; higher PRF allows higher velocities but shortens the usable depth.

Related: Aliasing, Doppler angle
Wall filter

A high-pass filter that removes low-frequency, high-amplitude Doppler signals arising from vessel wall motion and slow tissue movement, allowing the weaker signals from flowing blood to be displayed.

Related: Spectral broadening
Sample volume

The adjustable depth and length along the ultrasound beam from which pulsed Doppler echoes are processed, determining exactly where along the vessel the velocity measurement is obtained.

Related: Doppler angle, Spectral broadening
Doppler flow phantom

A test object containing fluid-filled vessels and blood-mimicking fluid pumped at known velocities, used to verify Doppler accuracy, sensitivity, and depth penetration during routine quality assurance.

Related: Quality assurance versus quality control
Sensitivity and specificity

Sensitivity is the proportion of diseased patients correctly identified by a test, while specificity is the proportion of healthy patients correctly identified as normal; both express accuracy against a reference standard.

Intersocietal Accreditation Commission (IAC)

A nonprofit organization that accredits vascular laboratories through peer review of case studies, personnel credentials, equipment, and quality assurance practices; accreditation is frequently required for reimbursement.

Related: Quality assurance versus quality control
Quality assurance versus quality control

Quality control covers day-to-day checks of equipment and test performance, while quality assurance is the broader program that includes protocols, personnel competency, correlation with outcomes, and continuous improvement.

Related: Intersocietal Accreditation Commission (IAC), Doppler flow phantom
ALARA principle

As Low As Reasonably Achievable: the practice of minimizing ultrasound exposure for patients and sonographers by limiting dwell time and output while still obtaining the diagnostic information needed.

Related: Quality assurance versus quality control
Thoracic outlet syndrome

Compression of the brachial plexus or subclavian vessels at the thoracic outlet, assessed with provocative maneuvers and duplex or plethysmographic monitoring of subclavian artery flow and waveform changes.

Related: Photoplethysmography (PPG), Subclavian steal syndrome

RVT study plan

8 weeks · About 7 hours per week

Week 1

Cerebrovascular & Instrumentation, Quality Assurance and Specialized Vascular Testing

  • Sketch the aortic arch, carotid bifurcation, and vertebral artery anatomy from memory, labeling ECA versus ICA differentiating features

  • Recite the velocity and spectral criteria for <50%, 50-69%, 70-99% ICA stenosis and complete occlusion without notes

  • Identify the technical pitfalls (Doppler angle, sample volume placement, wall filter, gain) that can falsely elevate or lower carotid velocities

  • Practice 30 minutes of hands-on or simulated knobology, changing scale, baseline, and color box to eliminate aliasing

  • Explain the hemodynamic rationale for parvus-tardus and tardus waveforms distal to a high-grade stenosis

Your checkpoint

A one-page handwritten carotid interpretation cheat sheet, plus a closed-book 20-item self-quiz scored at 85% or higher

Week 2

Cerebrovascular & Peripheral Arterial

  • Interpret vertebral artery spectral waveforms and describe the waveform sequence in subclavian steal

  • Contrast the expected duplex findings of carotid dissection, fibromuscular dysplasia, and carotid body tumor in writing

  • Summarize indications, limitations, and typical findings of transcranial Doppler in sickle cell screening and vasospasm monitoring

  • Classify lower-extremity arterial waveforms as triphasic, biphasic, or monophasic and match each to a physiologic cause

  • Compare normal, dampened, and absent Doppler signals across proximal, mid, and distal limb segments

Your checkpoint

Five written cerebrovascular case interpretations (history, findings, impression) graded against published criteria with at least 4 of 5 fully correct

Week 3

Peripheral Arterial

  • Calculate ankle-brachial index, toe-brachial index, and segmental pressure gradients from raw numbers and state the diagnostic category for each

  • Sequence the cuff placement and limb-position steps for a segmental pressure study and explain how to avoid cuff artifact

  • Describe the exercise and reactive hyperemia protocols, including indications and contraindications

  • State duplex velocity criteria for 50% and 70% peripheral arterial stenosis and identify when criteria do not apply (e.g., native versus graft, tandem lesions)

  • Outline a graft surveillance schedule and the duplex parameters that trigger intervention

Your checkpoint

A completed ABI/segmental pressure worksheet with 10 calculations solved at 100% accuracy, including a correct interpretation label for every case

Week 4

Abdominal and Visceral Vascular & Peripheral Arterial

  • Measure and document aortic diameter using outer-wall-to-outer-wall technique and state surveillance intervals by size

  • Interpret renal artery duplex using the renal-aortic ratio and direct velocity thresholds, and differentiate intrinsic from extrinsic compression

  • Evaluate celiac and SMA waveforms for stenosis, including the median arcuate ligament pattern

  • Explain portal vein waveform changes in portal hypertension and the expected findings after TIPS placement

  • Review the peripheral arterial material from weeks 2-3 and correct every error from the previous deliverable

Your checkpoint

A timed 50-item mixed midterm covering domains 1, 2, and 4, scored at 80% or higher, with a written list of the top three weak subtopics

Week 5

Abdominal and Visceral Vascular

  • Describe renal transplant waveforms, resistive index interpretation, and the signs of rejection, stenosis, and thrombosis

  • Detail mesenteric ischemia duplex findings, including the difference between acute and chronic presentations

  • Explain hemodialysis access surveillance: flow volume calculation, stenosis criteria, and the causes of access recirculation

  • Compare duplex appearance of true aneurysm, pseudoaneurysm, and endoleak following endovascular repair

  • Assemble a one-page summary of visceral and aortic normal values used in reporting

Your checkpoint

A 15-case abdominal and visceral vascular interpretation log with written impressions matched to a criteria sheet at 90% agreement

Week 6

Peripheral Venous

  • Sequence the lower-extremity venous duplex protocol, including patient positioning, augmentation maneuvers, and the calf vein approach

  • Differentiate acute from chronic deep vein thrombosis using vein caliber, echogenicity, collapsibility, and flow characteristics

  • Describe upper-extremity venous protocols, including effort thrombosis and central catheter-related evaluation

  • Interpret venous insufficiency studies by applying reflux time cutoffs for the great, small, and perforating veins and describing the provocation maneuvers

  • Outline vein mapping measurements required before bypass or ablation

Your checkpoint

Ten venous cases classified as normal, acute DVT, chronic DVT, or reflux with 90% or higher accuracy and a stated rationale for each

Week 7

Peripheral Venous & Instrumentation, Quality Assurance and Specialized Vascular Testing

  • Explain transducer selection, frequency trade-offs, and ALARA principles including thermal and mechanical indices

  • Describe routine quality assurance of the ultrasound system: phantom uniformity, sensitivity, axial and lateral resolution, and depth of penetration checks

  • List the elements of a vascular laboratory quality program including accreditation standards, credentialing, and reporting requirements

  • Explain specialized testing indications and findings for plethysmography, thoracic outlet maneuvers, cold sensitivity, and photoplethysmographic reflux

  • Review infection control, ergonomics, and patient safety practices relevant to vascular testing

Your checkpoint

A two-page instrumentation and QA review sheet covering ALARA, phantom testing, and one complete mock QA log entry

Week 8

Cerebrovascular & Peripheral Arterial & Peripheral Venous & Abdominal and Visceral Vascular & Instrumentation, Quality Assurance and Specialized Vascular Testing

  • Complete two full-length timed practice exams under exam-day conditions

  • Rank performance by domain and re-study the two lowest-scoring domains until scores reach 80% or higher

  • Rehearse written interpretation templates for the highest-frequency case types across all five domains

  • Practice single-best-answer strategy: eliminating distractors, identifying absolute words, and pacing 90 seconds per item

  • Consolidate all cheat sheets into a single final review document

Your checkpoint

Two full-length practice exams completed with an overall score of 85% or higher and no domain below 75%

Milestones

Week 2: Closed-book carotid criteria table reproduced with 100% of the 70% stenosis thresholds correct, and 20-item cerebrovascular self-quiz at 85% or higher

Week 4: Timed 50-item midterm across domains 1, 2, and 4 at 80% or higher with a written weak-area list

Week 6: Venous case set completed at 90% or higher, and midterm weak areas re-tested at 85% or higher

Week 8: Two full-length timed exams at 85% or higher overall and no domain below 75%

RVT exam day

Your checklist

Appointment confirmation and test-center address

Save a screenshot or printout of your appointment confirmation with the exact test-center address, date, and arrival time. Phone-based confirmations may not be accepted at every site, so a paper copy is the safer default.

Two acceptable forms of identification

Most testing centers require a primary government-issued photo ID with signature (driver's license or passport) and a secondary ID. Accepted ID combinations vary by provider and country, so confirm the exact list with the credentialing body before test day.

Name matching your registration exactly

The name on your ID should match the name on your exam registration. If you have recently married, changed a name, or use a variant spelling, resolve the discrepancy with the provider before the appointment date.

Arrival time buffer

Plan to arrive at least 30 minutes before your scheduled start to allow for check-in, ID verification, and secure-storage procedures. Confirm the recommended arrival window with your provider, since late arrivals may be refused entry.

Travel and parking plan

Scout the route in advance, note parking options or public transit times, and identify a backup route. Traffic and parking delays are the most common preventable cause of late arrival.

Knowledge of the on-screen tools and question format

Before test day, review the provider's tutorial or demo for navigation, flagging, review screens, and any supplied calculator or reference tools. Do not count on learning the interface during the exam.

A short, personal high-yield review sheet

Limit yourself to one or two pages of material you already know well, such as hemodynamic thresholds, normal velocity ranges, and waveform descriptors. This is for reinforcement only, not for learning new content on the morning of the exam.

Leave prohibited items at home or in the car

Phones, smartwatches, fitness trackers, notes, bags, and outerwear are typically not allowed in the testing room. Secure storage may be limited, so travel light.

Eyeglasses or contacts if you use them

You will be reading a screen for an extended period. Bring your usual corrective lenses and any needed medication, subject to the test center's rules on permitted personal items.

Pre-approved testing accommodations

If you need extra time, a separate room, or medical accommodations, these must be requested and approved well before the exam date. Same-day requests are generally not accommodated.

Rest and nutrition plan for the prior 48 hours

Protect sleep in the two nights before the exam and plan a normal, light meal for the morning of. Fatigue and low blood sugar degrade performance on long computer-based exams more than one extra hour of cramming helps.

Emergency contact and backup transport

Have a backup ride or transit option and the test center's phone number saved in case of a last-minute problem. If you cannot attend, contact the provider as early as possible to discuss rescheduling rules.

Plan the day

72 hours before

Reconfirm the appointment date, time, and test-center address. Verify your ID meets the provider's requirements and that your registered name matches. Finalize travel and parking logistics. Switch to review-only studying of material you already know, and stop introducing new topics.

72 to 48 hours before

Take one short practice set under timed conditions if you have not already, purely to rehearse pacing and the on-screen interface. Then stop adding new content. Protect both nights of sleep starting now rather than the night before.

24 hours before

Pack your ID, confirmation, glasses, and any permitted items. Lay out clothes and confirm your route and departure time. Do a brief pass over your one-page review sheet, then stop studying by early evening. Avoid alcohol and heavy meals, set two alarms, and go to bed at your normal time.

Morning of the exam

Eat a normal light meal and hydrate normally. Skip a long cram session; at most, skim your review sheet for 10 to 15 minutes. Arrive within the provider's recommended window, complete check-in and ID verification, and use the restroom before you are seated.

First minutes of the test

Read any on-screen instructions carefully, confirm the tools and navigation before beginning, and take a few slow breaths. Note the number of questions and any time markers so you can budget your pace.

During the test

Answer each item in one pass, flagging anything that needs more thought rather than stalling. Watch your time per question and leave room to revisit flagged items. If unscheduled breaks are permitted, keep them very short unless the clock is paused, and confirm that policy with the proctor beforehand.

Final minutes of the test

Return to flagged items, answer every question if the format does not penalize guessing, and review that no item was left blank. Submit only after confirming the review screen and any end-of-test prompts.

After the test

Note how and when results will be delivered, and keep your confirmation and score report for your records or credentialing file.

Rules to check

  • Bring identification that exactly matches your registered name, and confirm the accepted ID types with the credentialing body before test day; a mismatch can result in being turned away.

  • Expect a secure check-in: personal items such as phones, watches, and notes are typically stored outside the testing room, and accessing them during the exam is generally prohibited.

  • Break policies vary by provider and delivery format. Ask before you begin whether your clock stops during breaks, and plan your time accordingly rather than assuming.

  • Most certification exams are scored on a scaled basis rather than as a raw percentage, and results are usually reported as pass or fail. Do not try to infer a raw cut score from practice performance.

  • Exam content is confidential. Sharing, reproducing, or discussing specific exam items with others may violate the testing agreement you accept at check-in.

  • Accommodations and rescheduling must generally be arranged in advance through the provider. If an emergency arises, contact the provider as soon as possible to ask about your options.

  • Follow all proctor instructions, including rules on scratch materials, calculators, and movement in and out of the room. Irregular behavior can lead to score invalidation.

  • If you are unsure about any policy on ID, breaks, permitted items, or scoring, confirm it directly with the credentialing provider rather than relying on secondhand information.

RVT FAQ

How many RVT practice questions are available?

The practice set has 200 RVT questions. Each one includes the choices, the correct answer, and an explanation, and the set is free to read with no signup.

How many RVT flashcards are available?

The flashcard set has 32 RVT cards. Each card shows the term and its definition.

Which topics does the RVT study material cover?

The RVT study material has 5 sections, including Cerebrovascular, Peripheral Arterial, and Peripheral Venous.

What format facts does the catalog list for the RVT exam?

The catalog lists these item types: multiple choice for the RVT exam.