- Bremsstrahlung
- Braking radiation produced when high-speed electrons are deflected by tungsten nuclei; it forms the continuous x-ray spectrum and accounts for most dental photons.
- Characteristic radiation
- Produced when a projectile electron ejects an inner-shell tungsten electron and an outer electron fills the vacancy, emitting a photon of discrete energy.
- Inverse square law
- The intensity of the x-ray beam is inversely proportional to the square of the distance, so doubling the distance reduces intensity to one quarter.
- Half-value layer
- The thickness of aluminum that reduces x-ray beam intensity by half; it expresses beam quality and is used to verify adequate filtration.
- Photoelectric effect
- A photon transfers all its energy to an inner-shell electron and ejects it; dominates at low kVp in dense tissues such as enamel and bone, producing high contrast.
- Compton scatter
- A photon interacts with an outer-shell electron, loses part of its energy, and changes direction; this scatter increases with higher kVp and greater patient thickness.
- Occupational dose limit
- The annual effective dose limit for radiation workers is 50 mSv (5 rem); cumulative lifetime guidance is 10 mSv multiplied by age in years.
- Public dose limit
- The annual effective dose limit for the general public is 1 mSv (0.1 rem), with 0.5 mSv per month permitted for the fetus of a declared pregnant worker.
- Lead apron requirement
- Dental aprons must provide at least 0.25 mm lead equivalent and be inspected annually; store them flat or on a rack so cracks do not develop.
- Operator position
- During exposure the operator stands at least 6 feet (2 meters) from the tubehead and 90 to 135 degrees to the primary beam, never in its path.
- Rectangular collimation
- Restricts the beam to slightly larger than the receptor; it substantially reduces patient dose compared with a round PID opening.
- Dosimeter badge
- Worn at collar level outside the lead apron; it records cumulative dose and is exchanged and processed by an accredited monitoring laboratory monthly or quarterly.
- Step-up transformer
- Increases incoming line voltage to the high kilovoltage needed to accelerate electrons from the filament to the target inside the x-ray tube.
- Step-down transformer
- Reduces line voltage to about 3 to 5 volts, heating the tungsten filament so it releases electrons by thermionic emission.
- Position indicator device
- The open, lead-lined cylinder that aims and collimates the beam; 16-inch PIDs reduce divergence and lower patient skin dose.
- Aluminum filtration
- Removes low-energy, nonpenetrating photons from the beam; total filtration must be at least 1.5 mm aluminum equivalent on dental units up to 70 kVp.
- Focal spot
- The small tungsten target area where x-rays are produced; a smaller focal spot sharpens the image but limits the milliamperage that can be used.
- Milliamperage
- Controls the number of electrons flowing from filament to target and therefore the quantity of x-rays; with time it forms mAs, which governs receptor exposure.
- Kilovoltage peak
- Controls electron speed and beam penetrability; higher kVp produces a longer-scale, lower-contrast image with more visible gray levels.
- Fifteen percent rule
- A 15 percent increase in kVp doubles receptor exposure, so the mAs should be halved to keep image density unchanged.
- SLOB rule
- Same lingual, opposite buccal: an object moving the same direction as the tubehead is lingual, while opposite movement indicates buccal placement.
- Bisecting angle technique
- The receptor sits close to the teeth and the central ray is aimed perpendicular to the imaginary bisector; misalignment causes elongation or foreshortening.
- Standard precautions
- Treat all patients' blood, saliva, and body fluids as potentially infectious; the approach applies regardless of the patient's known medical status.
- Barrier technique
- Single-use plastic covers on the tubehead, PID, sensor, and controls; remove and replace them between patients and never reuse them.
- Sensor barrier
- Digital sensors and phosphor plates must be covered with disposable plastic sleeves; electronic sensors are never autoclaved or immersed in disinfectant.
- Hand hygiene
- Wash with soap and water or use an alcohol-based rub before and after glove use and after touching contaminated surfaces or items.
- Film packet handling
- Wipe contaminated film packets with disinfectant and dry them before processing; discard packets that are torn or heavily contaminated.
- Stepwedge test
- Exposed and processed weekly; density steps should remain uniform, revealing drift in developer strength, temperature, or processing time.
- Safelight test
- Expose a film partly covered by an opaque object, then process it; fogging indicates a light leak or an incorrect safelight filter or bulb.
- Retake log
- Records every repeated radiograph and its cause; reviewing it reveals technique or processing problems, with under 5 percent a common goal.
- Patient selection criteria
- Radiographs are ordered from clinical findings and caries risk rather than routinely; high-risk patients may need bitewings every 6 to 18 months.
- Equipment inspection
- Machine calibration, beam alignment, and lead apron checks are performed at least annually and after repairs, then documented in the quality assurance log.